Eutech EC 1720 Conductivity Meter 19
3.2.9 TDS Factor
Enter the TDS factor value used to calculate the TDS (total
dissolved solids) value shown in the main measure mode.
Press the
or
key to scroll through the Setup
Menu list until the TDS Factor item is shown.
Press the
or
key to set the TDF factor value.
a.
Press the
key to increase the value by 0.01.
Press and hold the
key to quickly increase.
b.
Press the
key to decrease the value by 0.01.
Press and hold the
key to quickly decrease.
Press the “Enter” key to save the change.
3.2.10 Read Type
Set the Read Type as Continuous, Auto-Read or Timed to
define how measurements are performed and when
measurements are saved and exported.
Continuous:
In the Continuous measure mode, measurements are
continuously updated on the display and the stopwatch
(stabilizing) or checkmark (stable) icon indicates the
measurement stability status. Press the “Log/Export” key
to save a measurement to the data log and export.
Auto-Read:
In the Auto-Read measure mode, press the “Measure” key
to start a measurement. When the measurement is stable,
the checkmark (stable) icon is shown and the
measurement is locked on the display until the “Measure”
key is pressed again. The stable measurement is
automatically saved to the data log and exported.
Timed:
In the Timed measure mode, the measurement values are
continuously updated on the display. Measurements are
automatically saved to the data log and exported at the
selected time intervals, from 5 seconds to 60 minutes, the
entire time the meter is in the measurement mode.
Press the
or
key to scroll through the Setup
Menu list until the Read Type item is shown.
Press the
or
key to set the read type.
Press the “Enter” key to save the change.
Summary of Contents for Eutech EC 1720
Page 2: ......