Note on the dose display:
The natural background radiation in Germany varies between 0.5 mSv and 2 mSv per year.
Thus, doses between 1 and 2 µSv per working day correspond to the background due to
natural environmental radiation.
Audible warnings:
The EPD delivers an audible warning (beep) on two occasions:
•
as an acknowledgment signal when the operating button is pressed
•
as a warning signal
Confirm warning signals: (hold for 2 seconds)
The EPD responds to certain events by delivering different warning signals, which can be
confirmed and muted with a long keystroke. During the warning, the reason for the warning
signals is indicated in the display screen and the LED signal (on the right side of the display
screen) is active. Both display screen and LED are blinking every second.
Standard configuration (surveillance of pregnant women):
Warning type
Display screen
Significance
Dose
10 (HP10-Alarm)
The total dose has exceeded
0,5 mSv (alarm1)
or 1 mSv (alarm2)
Dose rate
10 / h
The dose rate has risen above
mSv/h
The warning stops when the
dose rate falls below 0,9
mSv/h
Error
Fxxx
An error has occurred in the
unit
"Xxx" stands for the error
code
Misuse
Everything is blinking
Occurs when mechanical
vibrations are being
continuously recorded for 15
seconds
FO50KOM27A/ Guide for use