6
Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method
96
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Setting Up a HO Method
This section describes the dual measurement to determine hydrogen and oxygen from a single sample.
It contains the following topics:
•
•
•
“Creating a New Method for Dual Measurement”
•
“Creating a Sequence for Dual Measurement”
•
•
Procedure Overview
It is possible to perform dual measurements of hydrogen and oxygen from a single sample with the
system (Thermo Chemical Elemental Analyzer, ConFloIV and IRMS) in less than 7min offering fast
sample throughput and high productivity. The technique is suitable for both solid and liquid samples.
Different autosamplers are used to measure different sample types, see
.
For the analysis of two isotopic species (hydrogen and oxygen) from a single sample, a method that
comprises both must be defined. As soon as the hydrogen peak (
3
) has been identified,
Isodat Software Suite stops the HD acquisition. The magnet jumps to the CO configuration (defined
in the Switch Gas column in the time events list, see
).
HO
Figure 104.
Schematic Chromatogram of a HO Dual Measurement
Table 21.
Dual Measurement and Sample Type
Sample Type
Autosampler
Solids and viscous liquids (for example, honey,
serum, olive oil)
MAS 200, MAS Plus or others
Liquids, for example, water, alcohol, urine
AS3000, AS 1310, GC Pal, TriPlus 100 LS.