6
Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for H+O Measurement
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
105
4. In the
File New
dialog box, double-click
Sequence
. The
Sequence Properties
window opens, see
.
Figure 118.
Defining Number of Samples
5. In the
Sequence Properties
window, enter the
Number of Samples
, for example, 3.6.
6. Click
OK
. The sequence grid opens, see
Figure 119.
Creating a Sequence for Dual Measurement
7. For each row, enter the
Amount
of the individual sample, for example
0.260
.
8. Enable
to perform a
Peak Center
prior to measurement, see
9. Select an EA Method, for example
Py_iHeM.eam
.
10. Edit text in
Identifier 1
to identify the sample, for example benzoic acid.
11. Select
Method
, for example
Py_iHeM
.
12.
Save
the sequence with a name of your choice.
13. Continue with
Start Data Acquisition
To start the data acquisition
1. In the Sequence window, click
Start
to start the sequence.The
Start Sequence
window opens, see
.