Vacuum System and Gas Inlets
Vacuum System
Thermo Scientific
DSQ II Hardware Manual
155
Fore Pressure Gauge and Foreline Adapter
The fore pressure gauge and foreline adapter are connected to the high vacuum pump. The
fore pressure gauge measures the pressure in the foreline. If the pressure inside become too
high, the vacuum control board removes power to the high vacuum
pump to prevent damage.
The foreline adapter connects the foreline hose from the rotary-vane pump inlet to the high
vacuum
pump.
Figure 75.
Fore Pressure Gauge and Foreline Adapter (Right Side)
1
High Vacuum (Turbomolecular)
Pump
4
Fore Pressure Gauge
2
Foreline Adapter
5
High Vacuum (Turbomolecular)
Pump Outlet
3
Foreline Hose Outlet
2
1
3
5
4