Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
92
DSQ II Hardware Manual
Thermo Scientific
a.
Disconnect the foreline pressure gauge cable from the foreline pressure gauge.
b. Disconnect the foreline tubing from the gold fitting on the foreline adapter.
c.
Remove the foreline adapter clamp.
d. Remove the foreline hose clamp.
e.
Remove the current foreline pressure gauge and foreline adapter.
5. Install the new foreline pressure gauge (PN 119259-0002) and foreline adapter
(PN 119259-0003).
a.
Insert the new gauge and adapter.
b. Reattach the foreline hose and adapter clamps.
c.
Reconnect the foreline tubing.
d. Reconnect the cable.
6. Reattach the right, top, and front covers of the DSQ II.
7. Restart the DSQ II.