This message occurs if the flow rate of the sample stream being monitored by
the flow sensor drops below the lower limit.
To troubleshoot:
1. Verify the flow rate, using an external measuring device. If the flow rate is
normal, calibrate the flow sensor (see
2. If the message reappears, the sensor may need to be replaced. Contact
Technical Support for Dionex products for assistance.
3. If the problem persists, contact Technical Support for Dionex products for
assistance.
This error occurs if the offset value is out of the expected range when the flow
sensor is being calibrated.
To troubleshoot:
1. Review the calibration procedure and then rerun the calibration (see
2. If the problem persists, contact Technical Support for Dionex products for
assistance.
Flow sensor 1 flow rate is lower than limit
-or-
Flow sensor 2 flow rate is lower than limit
Flow sensor 1 offset calibration value is out of range
-or-
Flow sensor 2 offset calibration value is out of range
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...