3 • Operation and Maintenance
Doc. 065289-03 10/12
47
•
If the pressure is less than the expected amount, gas may be trapped in the
system. To release the gas, remove the pump fitting on the injection valve
port, labeled
P
(
2
). Allow the air to escape and then reconnect the fitting.
•
If the pressure fluctuates by more than about 0.13 MPa (20 psi), prime the
pump. See
for additional troubleshooting information.
•
If the pressure is too high, there may be a restriction in the system
plumbing. See
for troubleshooting information.
5. Verify that the baseline conductivity is at the expected reading for your
application and is stable. In general, it should be <30
S for a system set up
for anion analyses and <2
S for a system set up for cation analyses. See
for troubleshooting information if the conductivity is too high.
See
if the baseline is drifting or has excessive “noise” (large
fluctuations in readings).
6. Verify that the cell heater is at the set point and is stable. The temperature is at
equilibrium when the
Set Temperature
and
Current Temperature
readings
on the Control panel are the same.
7. Verify that the column heater temperature (if installed) is at the set point and
stable. The temperature is at equilibrium when the
Set Temperature
and
Current Temperature
readings on the Control panel are the same.
The system is now ready for sample processing.
3.9
Prepare Samples
NOTE Sample preparation can be performed while the system
is equilibrating.
3.9.1
Collecting and Storing Samples
Collect samples in high density polyethylene containers that have been
thoroughly cleaned with deionized (DI) water. Do not clean containers
with strong acids or detergents because these can leave traces of ions on
the container walls. The ions may interfere with the analysis.
If samples will not be analyzed on the day they are collected, filter them
through clean 0.45
m filters immediately after collection; otherwise,
bacteria in the samples may cause the ionic concentrations to change over
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...