Dionex Easion Operator’s Manual
88
Doc. 155028-01 08/20
A.8
Conductivity Detector and Flow Cell
Type
Bipolar-heated conductivity cell; microprocessor-controlled digital
signal processing over a range of 0 to 10,000
S/cm
Temperature
Compensation
Preset for accurate reading at 40 ºC
Cell Drive
8 kHz square wave
Auto Offset
-9,999 to 9,999
S
Linearity
1%
Resolution
0.0047 nS/cm
Calibration
Factory-calibrated before shipment to the user; calibrated in the
field with memory retention and transfer
Control and
Data Evaluation
Provided by Chromeleon software; communication with the
Dionex Easion is via USB (Universal Serial Bus)
Cell Body
PEEK
Cell Electrodes
Passivated 316 stainless steel
Cell Active
Volume
1
L
Maximum Cell
Operating
Pressure
10 MPa (1500 psi)
Cell
Temperature
Regulated at 40 ºC
Cell Chemical
Compatibility
Fully compatible with all typical ion chromatography eluents,
including MSA