B–3
Thermo Scientific Cryofuge 8 / 16 Centrifuges and Heavy Duty Centrifuges
Rotors
HAEMAFlex 6 Rotor
with Single Blood Bag Buckets (round)
Technical Data
Cryofuge 8
Centrifuge Voltage
200, 208, 220,
230, 240 V
380, 400, 415 V
Rotor Capacity
6 x 1000 ml
6 x 1000 ml
Maximum Speed n
max
5
000 rpm
5
000 rpm
Maximum RCF-Value at n
max
7
295 x g
7
295 x g
K-Factor at n
max
8
216
8
216
Radius max. / min.
261 mm / 116 mm
261 mm / 116 mm
Angle
90°
90°
Maximum Speed at 4 °C
5
000 rpm
5
000 rpm
Sample Temperature
at Max. Speed
(Ambient temperature of 23 °C,
run time 90 minutes)
50 Hz: 1 °C
60 Hz: -2 °C
50 Hz: 1 °C
60 Hz: -2 °C
Aerosol-tight
No
No
Rotor Body
Max. Cycle Number
50 000
Buckets
Max. Cycle Number
1
at
7295 x g / 5000 rpm
5000 x g / 4140 rpm
4000 x g / 3703 rpm
2500 x g / 2927 rpm
14 000
31 000
50 000
138 000
Weight
19.8 kg
Rotor Body (empty) with
Windshield Bowl
Windshield Lid
18.3 kg
1.5 kg
1
Cycle numbers stated are only valid, if the according speed (or rcf) is always set for centrifugation. If different speed (or rcf) settings are used, the
smallest cycle number stated is valid.
Table B–3:
HAEMAFlex 6 Rotor with Single Blood Bag Buckets (round)