5 Startup procedure
Thermo Scientific
12
50134805_03.2012
5 Startup procedure
DANGER!
Supply voltage and supply frequency must be within the range specified
for the power supply unit (3) or the control unit (9).
DANGER!
Magnetism.
Magnetic or metallic parts (e.g. data carriers, pacemakers, watches...)
can be affected by magnetic fields.
Keep such parts away from the magnetic stirrer (5) and the stirring
device (1).
DANGER
GLASS SPLINTER.
At high speeds of the mixing stick can lead to the destruction of the
sample vessels.
Wear safety goggles!
The vortex should not reach the stirring bar.
The stirring bar must rotate in the center of the sample vessel.
Watch the rotational behavior of the mixing stick for slow starts
CAUTION!
The equipment must not be operated in hazardous locations.
CAUTION!
Permitted ambient conditions for the directly controlled BIOSYSTEM:
Magnetic stirrer (5):
-10° to + 56°C at 100 % humidity.
Power supply unit (3):
+10° to + 40°C at max. 80 % humidity.
The magnetic stirrer (5) must not be operated in a water bath.
The power supply unit (3) must not be run in humid rooms, or set up in
water splash zones.
CAUTION!
Permitted ambient conditions for the remote-controlled BIOSYSTEM:
Magnetic stirrer (5): -10° to + 56°C at 100 % humidity,
0° to + 50°C in a water bath.
Control unit (9):
+10° to + 40°C at max. 80 % humidity.
The control unit (9) must not be run in humid rooms, or set up in water
splash zones.
CAUTION!
The magnetic stirring system must be switched off before connecting or
pulling the connector.
BIOSYSTEM DIRECT: The rotation speed control button (6) must always
be brought to position OFF first.
BIOSYSTEM: Switch off control unit (9) using power ON/OFF switch (18).