Operation - UT150 Controller
Using the Hold Function
Thermo Scientific
6-13
Ending a Program
To end a program while in the "RUN" at "HLD" (hold) mode, press and hold the
"UP/RESET" button to turn off the current program and extinguish the "RUN" or
"HLD" indicator.
Turning off the unit's power will also stop the program. When power is restored,
the controller is in the Reset or standby mode with no power to the heaters.
Using the Hold Function
To hold a running program:
1. Press and hold the SET/ENT key for 3 second, "noDE" will appear in the upper
display. Press SET/OUT key to "Hold".
2. Press the "arrow up" key so that "on" with flashing decimal appears in the
lower display.
3. Press the SET/ENT key to accept.
4. Press and hold the SET/ENT key again to return to the normal display.
To stop the hold mode:
Press and hold the SET/ENT key for 3 seconds. "noDE" will appear in the upper
display.Press SET/ENT key to "Hold". Press the "arrowdown" key so that "OFF"
with flashing decimal appears in the lower display.Press the SET/ENT key to
accept. Press and hold the SET/ENT key again to return to the normal display.
The programmer/controller will not operate the Unit's heaters (to
change or maintain a temperature) unless there is a program
running or a single setpoint value is selected in the Local Mode.
This Controller details to the "RESET" (rES) mode with each
power "ON". If power supply is introduced during program run or
single setpoint mode, these action operation mode must be
reselected.
Summary of Contents for CC58114PBC
Page 1: ......
Page 6: ...List of Figures ii Thermo Scientific ...
Page 12: ...Introduction 2 2 Thermo Scientific ...
Page 14: ...Unpacking 3 2 Thermo Scientific ...
Page 20: ...Installation Thermocouple Wiring Installation 4 6 Thermo Scientific ...
Page 22: ...Initial Start up 5 2 Heratherm Thermo Scientific ...
Page 46: ...Communication Option Addresses for Multiple Controllers 7 6 Thermo Scientific ...
Page 50: ...Excess Temperature Option Excess Temperature Controller Operation 8 4 Thermo Scientific ...
Page 57: ...Replacement Parts Thermo Scientific Heratherm 11 3 Figure 5 Wiring Diagram ...
Page 58: ...Replacement Parts 11 4 Heratherm Thermo Scientific ...
Page 59: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable ...
Page 60: ...Error Codes 12 2 Thermo Scientific ...
Page 62: ...WEEE Compliance 13 2 Thermo Scientific ...
Page 63: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable ...
Page 64: ...Spare Parts and Accessories 14 2 Thermo Scientific ...