Chapter 1
The Hybridization Oven 645
Unpacking and setup
8
GeneChip
™
Hybridization Oven 645 User Guide
1
Locating the oven
• Use a strong, level table or laboratory bench capable of supporting the weight of
the oven. The oven weigh
s
approximately 67 pounds (30 kg).
• Provide an adequate source of electrical power within reach of the power cord.
The oven operates on AC voltages from 100 to 240 volts nominal, 50/60 Hz, and
draws currents below 5 Amperes.
• Provide minimum clearance as follows: 0.5 inch (13 mm) from top, 2 inches (50
mm) from the back and each side, and 18 inches (460 mm) for door swing at the
front of the unit. Note that there is an air intake fan located in the rear panel. Do
not block this vent.
• Provide adequate bench clearance for cartridge preparation, loading and
unloading cartridge carriers.
• Avoid extreme ambient temperatures and locations close to heaters or air
conditioning vents.
• Keep the oven dry—avoid locations close to sinks or bulk liquid dispensers and
storage.
Unpacking the oven
NOTE:
The oven is packaged in a corrugated shipping crate that can be reuse if you
need to return the oven for service. However, if the crate has suffered significant
crushing, cutting or structural damage, and you need to return the oven to Thermo
Fisher Scientific, we recommend that you discard the crate and use new shipping
materials.
1.
Remove the oven from its corrugated box and styrofoam protectors.
2.
Place it on a sturdy table or laboratory bench.
3.
Remove protective wrappings.
4.
Pull on the handle to open the oven door and remove the box of accessories.
5.
Verify that the oven chamber is free of paper, cardboard or other debris.
6.
Push the door closed. It latches automatically.