Thermo Scientific Barnstead GenPure with xCAD
Ultrapure water system
48
12.3 Printer output
Various parameters are recorded by the printer. A differentiation is made between three types of
messages:
- Standard messages
- Code messages
- Fault messages
12.3.1 Standard messages:
Here a record of all measured values is printed out in dependence on the sending interval.
A complete data set is also printed out during NONSTOP-operation.
Print-out:
e.g.:
03.04.08 15:15
GenPure Standard
S.No. 3988/08
Interv. TC on UV off
LF1= 18.2 M
Ω
xcm
LF2= 10.0 M
Ω
xcm
LF3= 0.000 M
Ω
xcm
Temp.= 16.8 °C
TOC= 0 ppb
UV Intens.= 0%
The standard record includes all measured values. With systems that are not equipped
with TOC measurement and UV-intensity measurement, 0 is given as measured value!
12.3.2 Code messages:
Whenever a code number is entered in the system control and confirmed with the Enter-key,
then the code input is immediately printed out.
For code identification refer to the “Assignment Table for code numbers that allow the sys-
tem to be unlocked”.
Print-out:
e.g.:
03.04.08 15:17
GenPure Standard
S.No. 3988/08
Code 0002