Thermo Scientific Barnstead GenPure with xCAD
Ultrapure water system
39
During Rinsing the display shows:
Rinse
30 sec.
12.1.6 Disinfection
A sixth press on the menu-key gives you the possibility of carrying out disinfection. When
this menu-prompt is confirmed with Enter, the command
Disinfection cartidge. Install one
“
is shown. After confirming this with Enter, disinfection is carried out. The pump runs for the
disinfection time that has been set in the OEM-menu. When the half of the disinfection
time has passed, the rinsing solenoid valve opens and stays open up to the end of the dis-
infection process. The
"New filter cartridge. Install one"
message is given when the disin-
fection process has finished
On confirming this with Enter, the system returns to the last operating mode.
During disinfection, the count-down of the disinfection time is shown in the display
The display shows:
Disinfection
Press enter
After confirnation with Enter, the display shows:
Disinfection
cartridge!
Install one
Press enter
During disinfection, the display shows:
Disinfection
25 min.