Use a screwdriver to adjust the alignment screw
(#5) until the graph displays a maximum.
51
Tighten the thumbscrew.
52
Press
Esc
twice to return to the standard display.
53
If desired, reset the hours of lamp operation to zero
(see page 30).
54
Press
MODE
once to select Transmittance
mode.
55
Press
ZERO
once to zero the instrument and wait
until the display reads 100%T.
56
Using a screwdriver, gently move the tip of the
lamp slightly up and down. Watch the display to
find the position of maximum output.
57
Repeat steps #19 and #20. This final alignment
procedure compensates for lamp variability.
58
Replace the door of the lamp compartment and
allow the lamp to warm up for at least 30 minutes.
59
Use un destornillador Phillips para ajustar el
tornillo de alineación (#5) hasta que el gráfico
muestra un máximo.
60
Aprete el tornillo.
61
Oprima
Salir
dos veces para regresar a la pantalla
estándar.
62
Si desea, reajuste las horas de operación de la
lámpara a cero (vea pagina 30).
63
Oprima
MODE
una vez para seleccionar modo
Transmitancia .
64
Oprima
ZERO
una vez para encerar el
instrumento y espere hasta que la pantalla lea
100%T.
65
Usando un destornillador, mueva suavemente el
extremo de la lámpara. hacia arriba y abajo.
Observe la pantalla para encontrar la posición de
maxima salida.
66
Repita los pasos #19 y #20. Este procedimiento de
ajuste final compensa por variaciones en las
lámparas.
67
Reemplace la puerta del compartimiento de la
lámpara y permita calentar por espacio de 30
minutos.