Concentration measurements using a standard
1
Press
A/T/C
to select the
concentration mode. The current
mode appears on the display.
2
Press
nm
or
nm
to select the
wavelength.
Note:
Holding
either key will cause the
wavelength to change more
quickly.
3
Insert your blank into the cell holder
and close the sample door.
Note:
Position the cell so the light
(indicated by arrow in drawing)
passes through the clear walls.
4
Press
0 ABS/100%T
to zero the
blank, then remove the blank and
insert your standard.
5
Press the
Standard
soft key and
use the and keys to enter the
concentration of the standard, then
press the
Set C
soft key to calculate
and display the factor for the
selected standard.
6
Remove the standard and insert
your sample into the cell holder.
The calculated sample
concentration appears on the LCD
display.
Mediciones de Concentración usando un
estándar
1
Oprima
A/T/C
para seleccionar el
modo concentración. El modo elegido
aparece en pantalla.
2
Oprima
nm
or
nm
para
seleccionar la longitud de onda.
Nota:
Mantener
apretada la tecla hará que la
longitud de onda cambie más
rápido.
3
Inserte su blanco en el portaceldas y
cierre el compartimiento de
muestras.
Nota:
Ubique la celda de forma que la luz
(indicada por la flecha en el dibujo)
pase a través de las paredes claras.
4
Oprima
0 ABS/100%T
para llevar a
cero el blanco, luego remuevalo, e
inserte su estándar.
5
Oprima la tecla de función
Estándar
y
use las teclas y para entrar la
concentración del estándar, luego
oprima la tecla de función
Fijar C
para
calcular y ver el factor para el
estándar seleccionado.
6
Inserte
su
muestra en el portaceldas. La
concentración de la muestra aparece
en pantalla.