5.0

GUIDELINES FOR SOLVENT CHOICE

*The SPD units (SPD121P and SPD111V  w i t h a radiant cover) are better suited for limited intermittent
use of the solvents listed in Part II. Use of these solvents with the SPD111V (without a radiant cover)
and SPD101B may cause the lid material to become discolored and cause potential damage. For addi-
tional technical assistance with respect to your solvent choice, please contact the Applications Support
Specialists, at Thermo Savant at 1-800-634-8886 or 631-244-2929 or Fax 631-244-0606.

Part I Solvents suited for the SPD101B unit

Ethanol

Methanol

Water

Acetonitrile

Part II Solvents NOT suited for the SPD101B unit

Methylene Chloride

Chloroform

Ethyl Acetate

Hydrochloric Acid

Trifluoroacetic Acid (TFA)

Dimethyl Sulfoxide (DMSO)

Dimethyl Fluoride (DMF)

N-Methylpyrrolidinone (NMP)

Part III All other solvents not listed above

Call our Application Scientists @ 1-800-634-8886 or 631-244-2929 

to see if your solvent can be used in the SPD101B concentrator.

Summary of Contents for SPEEDVAC SPD Series

Page 1: ...SPEEDVAC CONCENTRATOR SPD SERIES BASIC SPD101B INSTRUCTION MANUAL A Thermo Electron business...

Page 2: ...red Table of Contents Section Section Title Overview Page 1 0 Overview of the SPD101B Unit 1 1 1 Installation of the SPD101B Unit 2 1 2 How to Hook Up the SPD101B Unit 2 Section Section Title Control...

Page 3: ...on 6 Right Side Panel 7 Chamber View Window 8 Vacuum Port 9 AC Socket 10 Main On Off Switch 11 Manual Cover Lock Release 12 Rear Access Panel 1 2 3 4 5 6 12 8 9 10 11 7 Figure 1 Front View Figure 2 Si...

Page 4: ...9 Fax 631 244 0606 3 If the unit is part of a larger system use vacuum tubing supplied to hook up unit See figure 4 4 Vacuum clamps must be put on vacuum ports of pump and SPD unit See item 8 on figur...

Page 5: ...view the preset parameters during the run 4 TEMPERATURE DISPLAY GREEN DISPLAY Indicates set temperature or actual temperature during pre heat or during a run from 45 C to 80 C or NO heat 5 RUN TIME D...

Page 6: ...p cover to be opened RED power light is ON The display lights up showing the default values TEMPERATURE 45 C GREEN TIME 0 00 HRS GREEN 3 Place sample tubes in rotor so load is balanced Secure rotor wi...

Page 7: ...ials 20 x 47mm v vials 1 dram vials 15 x 45mm 4ml 20 x 60mm v vials 18 x 52mm mini scintillation vials 28 x 60mm scintillation vials Microplates Deepwell Microplates 40 64 120 100 40 100 20 40 72 10 3...

Page 8: ...choice please contact the Applications Support Specialists at Thermo Savant at 1 800 634 8886 or 631 244 2929 or Fax 631 244 0606 Part I Solvents suited for the SPD101B unit Ethanol Methanol Water Ace...

Page 9: ...bsorbent towels Always wear gloves when cleaning and dispose of paper towel in appropriate designated refuse containers 3 Replace chamber seal if cracked Part number is 197 6020 00 4 Outside of unit c...

Page 10: ...equential damages Warranty work is subject to our inspection of the unit No instruments equipment or accessories will be accepted without a Return Material Authorization RMA number issued by Thermo Sa...

Page 11: ...YSTEM SpeedVac Concentrator SPD101B UVS400 Universal Vacuum Source ANT100 BV130 Bleeder Valve Vacuum Tubing COMPONENT HIGH VACUUM SYSTEM BV130 Bleeder Valve Flexible Vacuum Tubing DVG50 SpeedVac Conce...

Page 12: ...SPD SERIES SPEEDVAC FAMILY SPD101B SPD111V SPD121P...

Page 13: ......

Page 14: ...97 3000 00 Rev B 2000 Thermo Savant 100 Colin Drive Holbrook NY 11741 4306 USA Tel 631 244 2929 1 800 634 8886 Fax 631 244 0606 www thermosavant com E mail savantec savec com A Thermo Electron busines...