![Thermo orion 135A Instruction Manual Download Page 45](http://html1.mh-extra.com/html/thermo-orion/135a/135a_instruction-manual_1104439045.webp)
Models 135A and 136S Portable Conductivity Meters
Appendix
39
Specifications
Ranges
Conductivity:
0.1
m
S/cm to 1,000 mS/cm
(c > 0.8 cm
–1
)
0.1
m
S/cm to 500 mS/cm
(c = 0.2 to 0.8 cm
–1
)
0.01
m
S/cm to 199.9
m
S/cm
(c < 0.2 cm
–1
)
Temperature:
–20.0 to +120.0
°
C / –4 to 248
°
F
nLF: 0 to 120
°
C
Salinity:
0.0 to 45.0 g/kg (0 to 30
°
C)
TDS:
0 to 1,999 mg/l (10 to 40
°
C)
Display
LCD 35 x 67 mm, character height 15 mm
Measurement cycle Approx. 2 sec
Accuracy
(+ 1 count)
Conductivity:
< 0.5 % of measured value
1)
Temperature:
< 0.3 K
Input 1
(Sensor)
Multi-contact for 2 and 4-electrode sensors with inte-
grated temperature probe
Input 2
(Temperature)
4 mm sockets for separate temperature probe
Permissible cell
constant
0.010 to 199.9 cm
–1
(adjustable)
Sensor
standardization
Direct entry of the cell constants,
Automatic determination of the cell constants with
Thermo Orion conductivity standards
Sensor standardization with any known solutions
Meter self-test
During power-up, segment test, memory test, display
of model no. and software version
Temperature
measurement
NTC 30 k
W
(automatic recognition during power-up)
or manual temperature entry
Temperature
compensation
Linear characteristic: 0.01 to 9.99 %/
°
C
nLF (non-linear characteristic for ultrapure water and
natural water)
Data logger
Manual, interval-controlled or event-controlled
Data points
100 memory locations: conductivity, salinity or TDS,
with temperature, date and time
1)
For conductivities > 500 mS/cm < 1% meas. value
Summary of Contents for 135A
Page 36: ...Appendix 30...
Page 37: ...Models 135A and 136S Portable Conductivity Meters Appendix 31 Certificate of Conformity...
Page 38: ...Appendix 32...
Page 39: ...Models 135A and 136S Portable Conductivity Meters Appendix 33 Control Drawing...
Page 40: ...Appendix 34...
Page 54: ...Index 48 Time format configuring 10 W Warranty 41...
Page 55: ......