2.3
Power Configuration
The Micromax model numbers, and voltage and frequency
requirements are listed in the table below.
MODEL VOLTAGE
FREQUENCY
3590 120
60
Hz
3591
220 - 240
50/60 Hz
3592 (RF)
120
60 Hz
3593 (RF)
220 - 240
50 Hz
3594 (RF)
220,230
60 Hz
3595 100
50/60
3596 (RF)
100
50/60 Hz*
Ensure that your site is configured to match the
centrifuge's power requirements.
Plugging the Micromax
into incorrect voltage or frequency will void your
warranty.
* Line Frequency Selection
(3596, Micromax RF)
Locate the 50/60 Hz selector switch to the left of the power
receptacle at the rear of the centrifuge. Adjust this switch
to match the line frequency at the site.
Fuses
Fuses are located at the back of the unit.
To install fuses:
Locate the power entry module on the back side of the
unit. The removable fuse drawer is located in the
module. A small latch holds the drawer in place. Press
this latch, then slide the drawer out.
•
Cat. No. 3590: The fuse drawer will have one spare
and one active fuse installed at the factory. Fuse is
rated for 6.25 Amps (part no. 50606B).
•
Cat. No. 3591: The fuse drawer will have two active
fuses installed at the factory. Fuse is rated for 4.0
Amps (part no. 43689).
•
Cat. No. 3592: The fuse drawer will have one spare
and one active fuse installed at the factory. Fuse is
rated for 8 Amps (part no. 50606A).
Micromax Series
3
Service
Manual
Summary of Contents for Micromax 3590
Page 50: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 51: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 52: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 53: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 56: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 57: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 64: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 74: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 75: ......
Page 82: ......
Page 83: ......
Page 84: ......