Test
Use this procedure to verify operation of the braking
system:
WARNING:
This procedure requires operation of the unit with
power applied and the cabinet removed. Use caution
to avoid electric shock. See Section 7.1, Warnings and
Cautions.
1.
2.
3.
4.
5.
6.
7.
8.
Remove the lid and place the cabinet on its side,
leaving the display board connected.
Locate TP1 (GND) and J11, pin 3 (YEL) on the Power
PCB.
Verify DC braking voltage across these locations.
Expected measurement should be approximately 170
VDC.
Plug the unit in and verify that a properly loaded rotor is
installed.
Select an appropriate speed and time, and start a run.
During a run, this voltage should drop slightly (3 - 10
VDC).
If voltage is not present, verify resistance of the brake
resistor at J11 pins 1 (RED) and 3 (YEL). The
expected value is approximately 50
Ω
.
If the resistor is good, replace the Power PC board.
Micromax Series
42
Service
Manual
Summary of Contents for Micromax 3590
Page 50: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 51: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 52: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 53: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 56: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 57: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 64: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 74: ...ThermoIEC Technical Service Dept Printed 09 12 2002 Tel 781 449 8060 Fax 781 455 9749 ...
Page 75: ......
Page 82: ......
Page 83: ......
Page 84: ......