Environmental Chamber ___________________________________________________________________________Service
5 - 6
CUSTOMER:
JOB NUMBER:
UNIT SERIAL NUMBER:
CONTROL TYPE:
CO2
PREPARED BY:
SEn
MA
SP.hı
20.0
Ot 2
Pr.AL
hYS 2
0.3
rP
OFF
P.dec
0.0
Ftr.E
both
Ot 3
rMt
L9c2
ALO
LOC
0
1
Sc.LO
4.00
FLtr
2.0
AO3.U
MA
LAt2
nLAt
Sc.hı
20.00
Ot 1
hEAT
O3.Lo
4.00
SIL 2
On
PI.LE
-1.00
Ftb1
20.0
O3.Hi
20.00
dSP 2
On
PI.HE
21.00
PL 1
100.0
r3.50
Proc
Unıt
US
r9.LO
0.0
PSL 1
0.0
r3.Hi
0.0
1.Err
nLAt
r9.hı
20.0
PSh 1
100.0
r3.Lo
20.0
FAıL
OFF
SP.LO
0.0
nLF 1
OFF
r3.Co
0.0
dSP
nor
POht
0
2
CAL
0.0
3
rE.ht
0.15
A2Lo
0.0
4
A-M
Auto
htM
Pid
rA.ht
0.10
Aut
OFF
Pbht
1.0
A2hı
20.0
4
dSPL = Display Intensity 5 to 99%
rESt = Restore Factory Calibration (Select YES to restore)
Usrr = Restore User Settings (Select YES to restore user configuration)
USrS = Save User Settings (Select YES to save user configuration)
dFLt = Default Parameters (Select YES to reset to default configuration)
NOTE: Not all available parameters are listed on this page. See Watlow manual for more
information.
NOTES:
¹ - Lock out setting range 1 to 4. (Factory set at 3)
² - %power output display. Not configurable.
³ - User settable calibration offset.
ʍ
- User selectable alarm points.
WATLOW SD CONFIGURATION RECORD
8/2/2004
DATE:
1900226 & 1900227 KITS
FACTORY PAGE:
Press UP and DOWN arrow for six seconds.
GLS
SETUP PAGE:
Press UP and DOWN arrow for three seconds.
OPERATION PAGE:
Press ADVANCE key from home page. LOC = 0 to view and change.