Environmental Chamber__________________________________________________________________Routine Maintenance
4 - 1
Section 4 - Routine Maintenance
4.1 Cleaning the Incubator
De-energize all potential sources of energy to this
unit and lockout/tagout their controls. (O.S.H.A.
Regulation, Section 1910-147.)
The continued cleanliness of the stainless steel used in this
unit has a direct effect on the appearance and operation of the
unit. Use the mildest cleaning procedure that will do the job
effectively. Clean the outside of the incubator with soap and
water or with any non-abrasive commercial spray cleaner.
Clean the inside of the chamber with alcohol and/or soap and
water. Disinfect the interior panels with a general use laboratory
disinfectant, diluted according to the manufacturer’s instruc-
tions. Rinse the surface thoroughly after each cleaning and wipe
the surfaces dry. Always rub in the direction of the finish pol-
ish lines.
Do not use chlorinated solvents on stainless steel
as they can cause rusting and pitting.
Do not use volatile or aromatic solvents for clean-
ing inside the cabinet as their residue can contam-
inate the cabinet environment.
The Thermopane glass door may be cleaned with commer-
cial glass cleaner or with a solution of ammonia and water.