background image

8

www.thermaltronics.com   support email: [email protected]

CALIBRATION

Calibrating the Temperature

1. Change the voltage on the precision power supply until the display temperature comes within ±3°C of the defined 

temperature range as follows:

• If the display temperature is lower then the defined temperature then decrease the voltage value set on the 

precision power supply until the temperature is correct.

• If the display temperature is lower then the defined temperature then decrease the voltage value set on the 

precision power supply until the temperature is correct.

Note: Please make a note of this voltage value as it will be used later to calibrate the tester.
2. Use the  mini USB cable (ST-CABLE-1) to connect the TMT-ST10 to a computer through the Mini USB Port.
3. Run program commix.exe series port management program found on the installation cd
4. Select the correct settings as follows:

  • Baud Rate = 9600 

 

 

 

• DataBits = 8

  • Parity = None 

 

 

 

• Stop Bits = 1

  • Input = ASC 

 

 

 

• Show = ASC 

 

5. Select the correct COM Port and press “Open Port” to open the serial port.

• If the wrong port is select an error message “Open Port Failure” will appear.
• If the correct port is selected the “Open Port” button will change to “Close Port”.

6. Input the commands as follows and press the “Send” Button. 

Note: XX.XX is the set voltage from step 1

Example:

7. Your tip tester is now calibrated.

DataBits

Input ASC

Stop Bits

Parity

Show ASC

Baud Rate

Enter Commands here

Command

Command description

TMT-ST10 Response

#LINK\CR

Establish link to TMT-ST10

LINK OK\CR

#SET TmV=

XX.XX

\CR Calibrate the thermocouple voltage output to 

XX.XX 

(Must be in Temperature mode)

SET OK\CR

Command

Command description

TMT-ST10 Response

#LINK\CR

Establish link to TMT-ST10

LINK OK\CR

#SET TmV=

20.64

\CR Calibrate the thermocouple voltage output to 

20.64

 (Must be in Temperature mode)

SET OK\CR

Summary of Contents for TMT-ST10

Page 1: ...USER MANUAL TMT ST10 SOLDERING TIP TESTER www thermaltronics com ...

Page 2: ......

Page 3: ...ective within the warranty period Thermaltronics will repair or replace it free of charge at its sole option Warranty period is from the date of purchase by the original owner If the date of purchase cannot be substantiated the date of manufacture will be used as the start of the warranty period TABLE OF CONTENTS TMT ST10 SPECIFICATIONS 1 INTRODUCTION 1 SYSTEM FEATURES 2 CONTROL PANEL 3 UNPACKING ...

Page 4: ...digit Thermal Sensor K Type Thermocouple ST SNSR 1 Size W x H x D 85mm x 175mm x 48mm Weight 0 35 KG LCD display 48mm x 26 5mm USB port Mini USB port Data Storage 32 640 Events Certification Marks CE INTRODUCTION Congratulations on your purchase of the TMT ST10 solder tip tester This unit has been tested and inspected by Thermaltronics prior to shipment and with proper maintenance will give you ye...

Page 5: ...which the soldering iron is connected Note The tester must share the same ground as the soldering iron 5 Test Plate Touch the Soldering Tip on the test plate to begin testing for Tip to ground resistance ohm or Tip to ground voltage mV 6 Thermal Sensor To test temperature touch the thermal sensor with a soldering iron tip Do not push hard If the sensor is damaged or oxidized it can be changed by p...

Page 6: ...o zero Note Restarting the test meter will clear the calibration In Tip to Ground Leakage mode mV pressing RESET will reset a Current Test Data to zero Note Restarting the test meter will clear the calibration e Hold Button pressing the HOLD button will instruct the tester to display only the maximum test value Press HOLD again to release hold mode f Select Button pressing the SELECT button toggle...

Page 7: ...e one thermocouple on the 6 Thermal Sensor Ensure that the sensor polarity Red left and Blue right are not reversed 4 Optional Remove the Ground Wire from the box If voltage or resistance testing is needed connect one end to the 4 Ground Connection Socket Connect the other end to the power strip or socket to which the soldering iron is connected see System Features 5 Optional Remove the ST CABLE 1...

Page 8: ...To test Tip to Ground Resistance ohm place the solder tip onto the 5 Test Plate Allow a few seconds for the reading to stabilize 5 Click the SAVE button to save the tested value to b Tester Memory Operation Test Tip to Ground Leakage 1 Turn on the TMT ST10 tip tester 2 Press the SELECT button twice to toggle to the Tip to Ground Leakage mV test mode 3 Connect one end of the ground wire ST GWIRE to...

Page 9: ... downloaded to the computer Sync tip tester date and time after download if selected TMT ST10 date and time will be synched to the computer date and time Operation Retrieving Test Data 1 Turn on the TMT ST10 tip tester 2 Use the mini USB cable ST CABLE 1 to connect the TMT ST10 to a computer through the 3 Mini USB Port 3 Run TMT ST10 Data Output software 4 Note Only select 5 Options Clear tip test...

Page 10: ...d onto the TMT ST10 2 Turn on the TMT ST10 tip tester 3 Verify that the tip tester is in Temperature testing mode 4 Take off the ST SNSR 1 thermocouple from the TMT ST10 tip tester 5 Set the precision power supply output to V 20 64mV This voltage is the thermoelectric voltage of the K type thermocouple at 500 C 6 Connect the V 20 64 mV from the precision power supply to two connectors on the tip t...

Page 11: ...port management program found on the installation cd 4 Select the correct settings as follows Baud Rate 9600 DataBits 8 Parity None Stop Bits 1 Input ASC Show ASC 5 Select the correct COM Port and press Open Port to open the serial port If the wrong port is select an error message Open Port Failure will appear If the correct port is selected the Open Port button will change to Close Port 6 Input t...

Page 12: ...m ORDERING GUIDE TMT ST10 THERMALTRONICS SOLDER TIP TESTER PART NUMBER DESCRIPTION TMT ST10 Solder Tip Tester ST CABLE 1 Mini USB Cable ST SNSR 1 Temperature Sensors 10 Pack ST GWIRE 1 Ground Wire EU Plug ST GWIRE 2 Ground Wire US Plug ST GWIRE 3 Ground Wire CN Plug ...

Page 13: ...Support Email support thermaltronics com www thermaltronics com ...