TITAN-C SERIES USER MANUAL
31
CONTINUED FROM PREVIOUS TABLE
CONTINUES
Instruction
Camera Instructions Received
Note
Read the
automatic
correction
temperature
interval
Receive
AA 04 00 18 00 C6 EB AA
The return value of the
thermal camera is 1
byte, representing the
temperature interval for
automatic correction,
in units of 10 ℃. In the
example, the return value
0x14 represents 20/10 ℃
=2 ℃.
Return
55 05 00 18 33 14 B9 EB AA
Electronic
variable
times
Receive
Normal: AA 0D 00 2A 01 00 00 00
00 00 7F 02 FF 01 63 EB AA
1.1 times: AA 0D 00 2A 01 00 1D 00
17 00 61 02 E7 01 61 EB AA
1.2 times: AA 0D 00 2A 01 00 35 00
2B 00 49 02 D4 01 62 EB AA
1.3 times: AA 0D 00 2A 01 00 4A 00
3B 00 35 02 C3 01 62 EB AA
1.4 times: AA 0D 00 2A 01 00 5B 00
49 00 23 02 B5 01 61 EB AA
1.5 times: AA 0D 00 2A 01 01 6B 00
55 00 14 02 A9 01 63 EB AA
1.6 times: AA 0D 00 2A 01 01 78 00
60 00 06 02 9E 01 62 EB AA
1.7 times: AA 0D 00 2A 01 01 84 00
69 00 FB 01 95 01 62 EB AA
1.8 times: AA 0D 00 2A 01 01 8E 00
72 00 F0 01 8D 01 62 EB AA
1.9 times: AA 0D 00 2A 01 01 98 00
79 00 E7 01 85 01 62 EB AA
2.0 times: AA 0D 00 2A 01 01 A0 00
80 00 DF 01 7F 01 63 EB AA
2.1 times: AA 0D 00 2A 01 01 A8 00
86 00 D7 01 78 01 62 EB AA
2.2 times: AA 0D 00 2A 01 01 AF 00
8C 00 D0 01 73 01 63 EB AA
2.3 times: AA 0D 00 2A 01 01 B5 00
91 00 CA 01 6E 01 63 EB AA
2.4 times: AA 0D 00 2A 01 01 BB 00
95 00 C4 01 69 01 62 EB AA
2.5 times: AA 0D 00 2A 01 01 C0 00
9A 00 BF 01 65 01 63 EB AA
2.6 times: AA 0D 00 2A 01 01 C5 00
9E 00 BA 01 61 01 63 EB AA
2.7 times: AA 0D 00 2A 01 01 C9 00
A1 00 B5 01 5D 01 61 EB AA
Suppose the width of
detector array is W, the
height is H, and the
amplification factor of the
preset value is m(accurate
to 1 place after the decimal
point), then:
The upper left corner:
The lower right corner:
The left instruction takes
640x512 as an example.