3) START-UP INFORMATION
A detailed description of the individual system components.
The Power Pack (TLC-
2002
)
The TLC-
2002
Power Pack provides power to each TLC-2001 Laser Probe. The probes’ ports, numbered 1 to 4, are the
same electrical output. A TLC-
2001
Laser Probe should always be connected to location #1 first.
Figure 13. A detailed look at the Power Pack.
1. Remote Interlock Connector: The Remote Interlock Connector is capable of disconnecting the TLC-2001 Laser
Probe(s) from main power when removed. To remove, press the button on the connector and pull the Remote
Interlock Connector straight out. Main power is now disabled and no action will restore it until the Remote Interlock
Connector is replaced. To replace, insert straight into the matching receptacle.
2. Emergency Stop: This button shuts down the entire laser system immediately when pressed, disabling main power.
NOTE: Depressing the Red Emergency Stop Button immediately disables the laser system. Operation
may be resumed by rotating the Red Emergency Stop Button 1/4 of a turn to the right and
allowing it to “pop out” again.
3. Key Switch:
This is the main on / off switch for the laser system. When the key is turned to the left, the
system is off, disabling mains power. When the key is turned
¼
turn to the right, the system is on. The key can only be
removed when in the OFF position.
5
4
2
3
1