AMC-400 Series Digital Transmitter with Smart Sensor 2
46
<SEL >050
Select finish display
<PB >050
Enter Gasing
>GAS>ON
Gas On Instruction
Maximum 60seconds
PB or SEL to continue
>_uP
Increasing Concentration
PB or SEL to end
>dn_
Decreasing Concentration
PB or SEL to end
>2.00
Stable Concentration
(signal voltage)
Monitors for 10
consecutive stable
samples.
Minimum 25seconds.
PB or SEL to end
>GAS >OFF
Gas Off Instruction
SEL to Accept
PB or idle (Maximum
60) seconds to discard
(bad)
>Acc or bad Confirmation
1second
<SEL >-SC
SPAN Calibration Title
Exit to Span Gas Menu
<PB >A05
SPAN Calibration Choice
Scroll to next Choice
<SEL >A06
SERVICE Setup Choice
Select SERVICE Option
< PB >-SE
SERVICE Title
Scroll to Choices
<SEL >SE1
Service Mode1 Choice
Select Service Mode2
Setup
>co or no2
>000
Scrolling Gas Concentration
Display;
Displays Sensor Unit if >1
unit present, Sensor gas,
Sensor Gas Concentration.
Running Display. Unit
can be bump tested
without gas
concentration
reflected to monitor.
<PB or <SEL> or
10minutes
Exit Service Mode1
Exit Service Mode
<PB >SE1
Service Mode1 Choice
Scroll to next Choice
<SEL >SE2
Service Mode2 Choice
Select Service Mode2
Setup
<SEL >100
Default or Last Forced Gas
Concentration
Select to edit 1
st
position address
<SEL >
_
0
Select 2
nd
position
<SEL > 0
Select 3
rd
position