Immunity test
IEC60601 test level
Conducted RF
EN-IEC 61000-4-6 (2014)
3 Vrms + 6 Vrms (ISM +
Amateur)
Radiated RF
EN-IEC 61000-4-3 (2006) + A1 (2008) + A2 (2010)
3 V/m
Proximity fields from RF wireless communications equipment
EN-IEC 61000-4-3 (2006) + A1 (2008) + A2 (2010)
9-28 V/m
Electrical transient conduction along supply lines
ISO 7637-2 (2004)
Not applicable (system not
intended for use in vehicles)
10.2 Electromagnetic emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable (the device is suitable for use
in all establishments other than domestic and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes)
10.3 Recommended separation distances
Recommended separation distances between portable and mobile RF communications
equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the device as recommended below, according to the
maximum output power of the communications equipment.