User Manual / Technical Manual - Operation
26 of 49
5.2.5 Deaerate standard set with drip chamber
1.
Hold the disposable set in the upright position.
Make sure that the slide clamp (A) of the disposable
set is closed and the roller clamp (B) of the hospital
administration set is open.
2.
Hold the drip chamber (C) upside down.
3.
Remove the cap (D) from the T-connector (E) and connect
to a syringe.
4.
Remove excess air from the system.
5.
Disconnect the syringe from the T-connector (E) and
reconnect the cap.
6.
Put the drip chamber back in the drip chamber holder (F).
Warning!