Test
5. Turn the
navigation wheel
to set the source value to
20.0000 V
,
and then press the
navigation wheel
to
enter the selection and exit EDIT mode.
SrcA:+20.0000 V LimA:1.00000 A
AUTO
Source value set to 20.0000 V
Step 2: Set the source limit
1. Press the
LIMIT
key. You will see a blinking character
in the
LimA
value field.
2. Press the down
RANGE
key to select the 10 mA
limit range. Verify that the source limit value in the LimA
field is
10.0000 mA
.
SrcA:+20.0000 V LimA:10.0000mA
AUTO
Source limit set to 10.0000mA
Step 3: Set measurement function and range
1. Press the
MEAS
key until the V (voltage)
measurement function is selected. In the following figure,
the measurement function has been set to
V
.
SrcA:+20.0000 V LimA:10.0000mA
AUTO
Measurement function set to voltage (V)
2. Press the
AUTO
key as many times as needed
to select the AUTO range function. When AUTO is
selected, the 2601B-PULSE automatically selects the
best range for the measured value. You will briefly see
the display shown below, and then the main display
screen reappears.
SrcA:+20.0000 V LimA:10.0000mA
AUTO
AUTO