Setup – Pulsewidth
Pulse Width Configuration
• The selection of pulse width affects the dynamic range
and resolution of the measurement. With a narrow pulse
width there will be higher resolution and smaller
deadzone; however, the dynamic range will be
decreased. A wide pulse width will bring higher dynamic
range and measure comparatively longer distance, but
resolution and deadzone will be increased. Therefore,
users should make the choice between dynamic range
and deadzone. .
• The options for pulse width will change according to the
distance range selected.
• Under the parameter configuration menu use
and
to highlight “PulseWidth.” Press Enter to select a value.
(Press
to .exit.)
• “Auto” is the default setting.
• When the range is set to “Auto,” the pulse width
automatically is set to “Auto.”