From the ID-Tone pull-down menu (upper right) select the type of ID-Tone duration
required:
►
Standard ID-Tone
– 500ms of 1990Hz & 500ms of /- 2Hz
►
Fast ID-Tone
– 300ms of 1990Hz & 300ms of 3980Hz
►
Short-Low Tone
– 300ms of 1990Hz & 700ms of 3980Hz
►
Long-Low Tone
– 700ms of 1990Hz & 300ms of 3980Hz
Most users select the Standard ID-Tone for a consistent high tone warble.
To turn off the ID-tone select the tone duration from pull-down menu and select ID-
Tone OFF.
DVM Mode
To select DVM mode for measuring DC voltage across a pair, from the Composite
Clock (T1) pull-down menu (upper left) select DMV mode. Connect a bantam jack to
the
SINGLE port jack
of the test set and the other end (alligators) to the Tip & Ring of
the pair you want to measure DC voltage. Also, so 60Hz AC does not affect the voltage
readings connect Frame Ground to the white banana GND connector of the test set.
The test set measures DC volts from 3V to 58VDC +/- 1Vdc.
To exit DVM simply select Composite or T1 from the DVM pull-down menu.
Summary of Contents for Greenlee Communications DataScout 10G
Page 1: ...2015 Greenlee Communications 52071568 Users Manual DataScout 10G 10Gx Version 11...
Page 2: ......
Page 4: ......
Page 12: ......
Page 16: ......
Page 22: ......
Page 26: ......
Page 40: ......
Page 58: ......
Page 66: ......
Page 86: ......
Page 102: ......
Page 114: ......
Page 115: ......