background image

Texmate, Inc. Tel. (760) 598-9899  •  www.texmate.com

FX-B101Q-DCA_DCV manual (d0103)

Page 5

STeP A  ReVIeW THe INPuT mODule STATuS

     1)  See pages 10 – 15 for information on input modules that may be used with this meter.

     2)  Confirm that the correct range and input is selected on the input signal conditioning 

module.

Note: When undertaking an initial set up and primary scaling and calibration of the meter it is 

best to start with a reset of the scaling.

STeP B  ReSeT THe SCAlINg

  1)  Apply power to the meter and press the ZERO and  SPAN buttons simultaneously for 2 

seconds. This  erases  any previously  memorized  scalings,  and  resets the scaling  to the 

factory default, of approximately zero to full scale, for an input, that is 0 to 100% of the range 

selected on the input signal conditioner.

Standard or Center Zero Display mode may be selected, depending on the Operating Mode select-

ed. If the standard display mode is not already selected open the meter case as showing on page 4 

and move the jumper clips on the display mode select header to the OFF position.

Standard Display with 

Jumper Clips in OFF

 

position

Standard Display Mode Calibration Procedure

100

0

Zero

Span

Lo

Hi

ANALOG OUTPUT

Reset the scaling 

to the default value

on by pressing

the Zero and

Span buttons

simultaneously

for 2 secs.

1

Z

C

3

2

4

Note: To calibrate the bargraph you must be able to input two input signals. Usually the minimum input (LO Input) and the maximum 

input  (HI Input) signals are used for optimum accuracy. However a scaling can be accomplished with any two signals that are higher 

or lower than each other by more than 5% of fullscale and are not greater than +2.1VDC or less than -1.05VDC

 

STeP C  SeT THe lOW INPuT SIgNAl ReADINg ON THe BAR

 

1)  Apply the LO input signal (4ma in this example) to the input pins.

 

2)  Using the ZERO button adjust the bar down to the required position. 

STeP D  SeT THe HIgH INPuT SIgNAl ReADINg ON THe BAR

 

1)  Apply the high input signal (20mA in this example) to the input pins. 

 

Using the SPAN button adjust the bar to the required position. 

 

This position could be higher or lower than the position adjusted in 

 

Step 2. The scaling for an input of 4 to 20mA is now complete.

Apply 4 mA 

to the 

Input Pins 

and adjust

bar display 

to the required 

position

100

0

Zero

Span

Lo

Hi

ANALOG OUTPUT

Apply 20 mA 

to the

Input Pins 

and adjust

bar display 

to the required 

position

100

0

Zero

Span

Lo

Hi

ANALOG OUTPUT

Apply 0 mA 

to the Input Pins 

and adjust

bar display 

to the required 

position

100

0

Zero

Span

Lo

Hi

ANALOG OUTPUT

One Point Quickset Rescaling and Calibration Procedure

ONe POINT ReCAlIBRATION

As explained earlier, the FX-B101Q bargraph is calibrated using two point calibration. Once a bargraph is calibrated, the low end of the 

range may be then recalibrated without affecting the calibration of the high end, and vice versa.

For example, take an FX-B101Q that has been calibrated to read zero to full scale for an input of 4 to 20mA. If now the scaling has to be 

changed to read zero to full scale for an input of 0 to 20mA, only the low (4 mA) end needs to be recalibrated. The high (20 mA) end of 

the scaling is left untouched, and so does not change. The following one point recalibration procedure is used for this purpose.

STeP A  ReCAlIBRATe THe lOW INPuT SIgNAl ReADINg ON THe BAR

 

1)  Apply the LO input signal (0ma in this example) to the input pins. 

 

The first segment will flash, indicating an under range condition.

 

2)  Using the ZERO button adjust the bar up to the required position.

 

3)  The FX-B101Q has now been recalibrated to read zero to fullscale 

 

for a 0 to 20 mA input.

Standard Display Mode Calibration Procedure

Standard Display Mode Calibration Procedure

Two Point Quickset Scaling and Calibration Procedure (continued)

Summary of Contents for LYNX FX-B101Q-DCA

Page 1: ...Center zero setting header selectable Provision for external brightness setting switch by connect ing the DIM to the GND pin on the back of the meter Smart averaging to speed up display response Opti...

Page 2: ...WN Program Direction Indicator 101 Segment Bargraph Controls and Indicators Controls and Indicators Controls and Indicators Glossary of Programming Symbols and Modes of Operation Standard or Center Ze...

Page 3: ...es to be pressed 3 While the bar is being adjusted a new offset and scale factor is continuously being calculated At the moment the button is released and the scaling is accepted the calculation data...

Page 4: ...ezel lightly lever the plastic catches up and for ward in the sequence shown by the num bers To remove rear cover plate press down lightly to release two plastic catches on either side of the case and...

Page 5: ...e to the input pins 2 Using the ZERO button adjust the bar down to the required position STEP D Set the High Input Signal Reading on the Bar 1 Apply the high input signal 20mA in this example to the i...

Page 6: ...is example When the HI button is pressed the UP or DOWN indicator LED shows the direction of change Release the HI button and press again to reverse the direction of change Initially the output change...

Page 7: ...in the meter s programmed parameters can be viewed but not changed Pin 20 COMMON To activate the LOCK or DIM functions from the rear of the meter the respective pins have to be connected to the COMMON...

Page 8: ...t Zero Offset Pot Input Range Header Input Range Header Input Range Header ID01 DC Volts 2 20 200V Custom w 24V DC Exc Custom 200V 20V 2V ON OFF 24V Exc 24V Exc Decrease Span Increase SPAN DC VOLTS PI...

Page 9: ...to a Hi and a Lo range This has the effect of dividing the adjustment range of the SPAN pot into ten equal 10 steps across 100 of the input Signal Span Range HI LO HI LO SPAN Turn Clockwise to Increas...

Page 10: ...ith Co Logo 170 ART FS1 Small run custom Faceplate 1 color 40 ART FS2 Small run custom Faceplate 2 color 45 ART FS3 Small run custom Faceplate 3 color 54 ART FS4 Small run custom Faceplate 4 color 60...

Page 11: ...ay Texmate s panel adapter enables modern DIN meters to fit in existing cutouts individually or stacked when replacing old 6 edgewise mechanical pointer meters Fits existing cut outs for 6 150 mm edge...

Page 12: ...snaps on the 36x144 mm 1 42 x5 69 case and enables single unit or stack mounting in an existing 6 edgewise pointer meter cut out 100 0 25 50 75 100 0 25 50 75 SP1 SP2 SP3 SP4 Zero Span Zero Span Pane...

Page 13: ...4X splash proof 30 OP PA 144X36 Panel Adapter for 144x36mm from 6 inch cutout 10 For Custom Face Plates see page 10 Prices subject to change without notice WARRANTY Texmate warrants that its products...

Reviews: