background image

Texmate, Inc. Tel. (760) 598-9899  •  www.texmate.com

FX-B101Q-DCA_DCV manual (d0103)

Page 3

Input Hi

Input Low

std

Input Hi

Input Low

std Inv

FLB202Q MODE O1

+ Input 

- Input 

+ Input 

- Input 

+ Input 

- Input 

0

0

0

biploar center zero

Input Hi

Input Hi

Input Hi

Input Low Input Low

Input Low

center zer

o

1/2

1/2

1/2

Standard Scaling 

Standard display mode selected and 

scaled so bar increases as input signal 

increases from Low to Hi.

Halfway Zero Point 

Center point display mode selected and 

scaled, so the bar increases upwards or 

downwards from the center point, for sig-

nals that are greater or less than half the 

 

calibrated full scale range respectively. 

When the input is equal to half the full 

scale range, only the center segment will 

be on.

Inverse Scaling

Standard display mode selected and 

scaled so the bar increases as the input 

signal decreases from Hi to Low.

Bipolar Center Zero 

Center point display mode selected and 

scaled, so the bar increases upward from 

zero, for increasing positive inputs and 

downward from zero for increasing negative 

inputs. When the input is zero, only the cen-

ter segment will be on.

Standard Display Mode

Center Zero Display Mode

Over View of Display Modes, Scaling Capabilities and Operating Modes  

1

3

2

4

Z

C

1

Z

C

3

2

4

 Meters with QuICkSeT PROgRAmmINg feature a unique, easy-to-use, two point scaling and calibration system.

Scaling or calibration is accomplished simply, by applying a zero or low input signal and adjusting the bar to the desired reading, using the 

ZERO button. A higher input signal is then applied, and the bar is adjusted to the desired reading for that input value, using the SPAN button.

ImPORTANT DeTAIlS THAT mAke QuICkSeT PROgRAmmINg eASy TO uSe AND uNDeRSTAND

1.  The zero and span buttons are functionally the same, except as follows: The ZERO button can initiate a scaling with input signals from 

zero to 95% of fullscale.The Span button can initiate a scaling with input signals from 5% of fullscale to 105% of fullscale.

2.  When a Zero or Span button is pressed, the Up or Down indicator LED will immediately light up to show the direction, in which the Bar 

will move, after a 0.5 second delay. If the button is released and pressed again, the opposite 

u

p or 

D

own indicator will light up, and 0.5 

seconds later the Bar will begin to move in that direction until the button is released. When the bar is being adjusted to zero or fullscale, 

the bar will automatically stop at the zero or fullscale position, and will not overshoot these positions, even if the button continues to be 

pressed.

3.  While the bar is being adjusted, a new offset and scale factor is continuously being calculated. At the moment the button is released, 

and the scaling is accepted, the calculation data is memorized and implemented. The Scaling calculation is based on the new position 

of the Bar, the input signal being applied at that moment, and the previously memorized position of the Bar and the input signal that was 

being applied, when the other button was last released.

4.  Positive and negative signals maybe integrated into a two point scaling. However when either a ZERO or SPAN button is pressed the 

input signal being applied, must be more than 5% higher or lower than the previously memorized value of the input signal, that was 

being applied when the other button was last released. If not, the bar will flash, the scaling will not be accepted, and the previous scaling 

will still be retained in memory.

5.  Because of the requirement, that a new scaling input signal must be 5% higher or lower than the previously stored value, it can some-

times be difficult to implement a desired scaling, particularly when using a calibrator that only has fixed output values. In this case Reset 

the Scaling by pressing the ZERO and SPAN buttons simultaneously for two seconds. Both scaling memories will be erased and an 

internal default scale factor will be loaded. This provides a display of zero to fullscale on the bar for an input of approximately 0 to 100% 

of the range selected on the input signal conditioning module. After Resetting the Scaling a new calibration, using either button, can be 

implemented with new input signal values. It is good practice to always use the Zero button for lower input signals and the Span button 

for higher input signals, even when the bar display scale is inversed.

6.  The larger the difference between two points used for calibration, the better the accuracy. However if the difference is too high, and 

the output from the input signal conditioning module is greater than +2.1VDC, or less than -1.05VDC, the bar will flash over range. The 

calibration will not then be accepted and, the previous scaling will still be retained in memory.  In this case, either a lower input signal 

must be used, or a higher range on the input module should be selected to recalibrate the meter.

Note: Most input signal conditioners have provisions for analog calibration and scaling. If the meterʼs scale factor is set to read zero with a 

zero input (shorted input), and to read 10 Bars fullscale with a 2.000 V input, any pre-calibrated signal conditioner with an output that does 

not exceed – 1 V to + 2 V, will read correctly in the meter without any further calibration.

Two Point Quickset Scaling and Calibration

Summary of Contents for LYNX FX-B101Q-DCA

Page 1: ...Center zero setting header selectable Provision for external brightness setting switch by connect ing the DIM to the GND pin on the back of the meter Smart averaging to speed up display response Opti...

Page 2: ...WN Program Direction Indicator 101 Segment Bargraph Controls and Indicators Controls and Indicators Controls and Indicators Glossary of Programming Symbols and Modes of Operation Standard or Center Ze...

Page 3: ...es to be pressed 3 While the bar is being adjusted a new offset and scale factor is continuously being calculated At the moment the button is released and the scaling is accepted the calculation data...

Page 4: ...ezel lightly lever the plastic catches up and for ward in the sequence shown by the num bers To remove rear cover plate press down lightly to release two plastic catches on either side of the case and...

Page 5: ...e to the input pins 2 Using the ZERO button adjust the bar down to the required position STEP D Set the High Input Signal Reading on the Bar 1 Apply the high input signal 20mA in this example to the i...

Page 6: ...is example When the HI button is pressed the UP or DOWN indicator LED shows the direction of change Release the HI button and press again to reverse the direction of change Initially the output change...

Page 7: ...in the meter s programmed parameters can be viewed but not changed Pin 20 COMMON To activate the LOCK or DIM functions from the rear of the meter the respective pins have to be connected to the COMMON...

Page 8: ...t Zero Offset Pot Input Range Header Input Range Header Input Range Header ID01 DC Volts 2 20 200V Custom w 24V DC Exc Custom 200V 20V 2V ON OFF 24V Exc 24V Exc Decrease Span Increase SPAN DC VOLTS PI...

Page 9: ...to a Hi and a Lo range This has the effect of dividing the adjustment range of the SPAN pot into ten equal 10 steps across 100 of the input Signal Span Range HI LO HI LO SPAN Turn Clockwise to Increas...

Page 10: ...ith Co Logo 170 ART FS1 Small run custom Faceplate 1 color 40 ART FS2 Small run custom Faceplate 2 color 45 ART FS3 Small run custom Faceplate 3 color 54 ART FS4 Small run custom Faceplate 4 color 60...

Page 11: ...ay Texmate s panel adapter enables modern DIN meters to fit in existing cutouts individually or stacked when replacing old 6 edgewise mechanical pointer meters Fits existing cut outs for 6 150 mm edge...

Page 12: ...snaps on the 36x144 mm 1 42 x5 69 case and enables single unit or stack mounting in an existing 6 edgewise pointer meter cut out 100 0 25 50 75 100 0 25 50 75 SP1 SP2 SP3 SP4 Zero Span Zero Span Pane...

Page 13: ...4X splash proof 30 OP PA 144X36 Panel Adapter for 144x36mm from 6 inch cutout 10 For Custom Face Plates see page 10 Prices subject to change without notice WARRANTY Texmate warrants that its products...

Reviews: