35
6-2. Application: Compare measurement
Applicable to
( )
+
DCV
ACI
DCI
2/4W
ACV
Hz/P
TEMP
( )
+
Background
Compare measurement checks and updates if the measurement data stays
between the upper (high) and lower (low) limit specified.
1. Activate Compare
measurement
Press the Shift key, then the HOLD (Comp) key.
2. High limit setting
1st display
Shows the high limit value
2nd display
Indicates high limit setting
1. Use the Left/Right key to move the cursor
(flashing point) between high/low setting, digits,
and decimal point.
2.
Change the parameter using the Up/Down key.
3.
Press the ENTER key to confirm editing and
move to the low limit setting.
3. Low limit setting
1st display
Shows the low limit value
2nd display
Indicates low limit setting
Set the low limit in the same way as in the high limit.
Press the ENTER key to confirm editing. The
compare measurement starts right away.
4. Compare
measurement appears
COMP
Indicates Compare mode
2
nd
display
Shows the compare measurement result: Pass, High, or Low.
5. Result
High
If the 2
nd
display shows High, the
result is above the High limit.
Digital I/O: FAIL Out (Pin 6) and HIGH Limit FAIL Out (Pin 7)
are activated.
Low
If the 2
nd
display shows Low, the
result is below the Low limit.