Software Overview
42
SBOU123A – March 2012 – Revised September 2016
Copyright © 2012–2016, Texas Instruments Incorporated
XTR108EVM-USB Evaluation Board and Software Tutorial
5.4.6
Verify Calibration
If the percentage of minimum-scale, mid-scale, and maximum-scale error after calibration is less than or
equal to 0.1%, the
Status
indicator displays the message
Calibration successful! Error < 0.1%
and the
square status LED turns on (bright green).
shows an example of a successful post-calibration
screen.
If the percentage of error after calibration is greater than 0.1%, the
Status
indicator displays the message
Calibration unsuccessful. Error > 0.1%
and the square status LED turns on (red).
Figure 28. Automatic Calibration Tab, Post-Calibration