B USB2ANY Firmware Upgrade
When the onboard USB2ANY programmer is first connected, or if the firmware revision used for the onboard
USB2ANY programmer does not match the version used by TICS Pro (2.7.0.0), TICS Pro will request a firmware
update. Follow the pop-up instructions to complete the update.
1. When the
USB2ANY Firmware Requirement
pop-up window appears, click
OK
to continue.
Figure B-1. Firmware Requirement
2. The
pop-up window will load. Disconnect the USB cable from the EVM.
Figure B-2. Firmware Loader
3. While pressing the BSL button (shown below), connect the USB2ANY cable.
Figure B-3. BSL Button Location
USB2ANY Firmware Upgrade
SNAU266 – JULY 2021
LMX1204EVM User's Guide
25
Copyright © 2021 Texas Instruments Incorporated