Figure 3-2. Bypass Mode Settings
2. Once these settings are verified, press the
Device Bringup
button.
3. Wait until after the message
Device Bringup Completed
appears in the Log.
4. Under the
Analog and Clock
tab, toggle
Dither Enable
to the
off
position.
5. In HSDC Pro, press the
Capture
button. You should now see something similar to
.
6. Adjust signal generator output such that the measured fundamental power in HSDC Pro is at the user's
desired level. (shown below is -12 dBFs)
Figure 3-3. 300.039 MHz, bypass mode, 2x averaging, dither off
Quick Start Procedure for Bypass Mode
SBAU377 – SEPTEMBER 2021
ADC32RF5xEVM User Guide
7
Copyright © 2021 Texas Instruments Incorporated