testo 6600
- 3.3 Using the software
137
7
Confirm control query to perform the reset.
The values are reset to the default settings.
8
Click on
[OK]
or
[Cancel]
to close the dialogue.
3.3.4 Adjusting the transmitter
This function is used to adjust an attached instrument. The following
adjustments may be carried out using the software:
1-point adjustment (offset)
2-point adjustment (upper and lower adjustment point)
Analog adjustment (entry via assistant/wizard).
See also
volume 1, chapter 1.3.5
Adjusting the instrument
, page 53
).
3.3.4.1 1-point
adjustment
The testo 400/650 with precision humidity probe (order no.
0636 9741) is recommended as the reference measuring
instrument for 1-point adjustment (offset) (see
volume 1,
chapter 1.3.5.2, page 55
).
1
Expose the reference measuring instrument and the instrument to be
adjusted to the same constant conditions and wait for equalization
period to lapse.
2
Mark the instrument file of the connected instrument.
3
Click on
[Adjusting the transmitter]
.
The
Properties of <Instrument type> <Serial number>
dialogue is
opened with the
Adjusting the transmitter
register.
4
Enter reference value and click on
[Carry out 1-point adjustment]
.
5
Confirm confirmation request.
The adjustment is carried out.