background image

7

M. 00

N. 0000

→ 

The symbols on the top line are explained below

Name of input socket and parameter

→ 

Displays reading in line 1

Name of location

Time/number of points in mean calculation

Displays measurement functions

Explanation of symbols:

Counter for the log number in the memory.
When saving manually: number of the measurement saved. 
When saving automatically: number of the measurement series. This
counter is needed in order to be able to find single logs or a measurement
series when reading out the memory.

Counter for saving a measurement cycle (required only for automatic
saving). The measurement cycle in a measurement series can be found.

Manual saving of a single measurement by pressing the save button 

.

Automatic saving program has been set up. 
Saving is activated by pressing the 

button.

Symbol for reading contents of memory on display.

Symbol for deleting memory contents.

If this symbol appears, the printing function is activated.
The symbol flashes while data is being transmitted. You can print on the
desktop printer by pressing the print button

.

Shows capacity of battery and rechargeable battery. 

If the inner segment no longer appears (symbol flashes), the battery has to
be changed or the rechargeable battery has to be recharged. The
instrument switches itself off automatically after 1 minute. 

Instrument description

Display

1.888.610.7664 

[email protected]

www.

Testo-Direct

.com

Summary of Contents for 545

Page 1: ...Instruction manual en testo 545 Luminous intensity measuring instrument 1 888 610 7664 info Testo Direct com www Testo Direct com ...

Page 2: ... 11 Freezing readings 11 Maximum readings 11 Minimum readings 11 Multi point mean calculation 12 Timed mean calculation 12 Location selection 14 Measuring range switchover function 14 Memory settings 15 Overview 15 Manual Automatic saving 16 Reading out or printing memory 17 Clearing memory contents Sample printouts 18 1999 Copyright Testo GmbH Co The software and software structure included in th...

Page 3: ...configuration 19 Power save function 19 Setting date time 19 Unit selection Factory reset 20 Power supply 20 Error messages 21 Technical data 22 Ordering data 23 3 1 888 610 7664 info Testo Direct com www Testo Direct com ...

Page 4: ...ommand is not possible if the PC cable is connected Opening the instrument inexpert handling and use of force cancels your warranty Putting in the batteries 9V block battery is included in delivery Open the battery compartment at the back of the instrument Put in block battery Observe polarisation Close battery compartment Refer to Power supply Chapter for more information on alternative power sup...

Page 5: ...er you will still need to update or define the data in the instrument Date Time Auto Off Units Some things can only be set via PC software See Ordering data Location name 8 characters Log heading 24 characters e g your company name is printed when the readings are printed i Instrument configuration First measurement 1 888 610 7664 info Testo Direct com www Testo Direct com ...

Page 6: ...ection assignment Prints On Off Saves Hold Max Min Mean Arrow buttons move the cursor Return to current measurement location selection Confirms menu setting executes function Keypad 1 888 610 7664 info Testo Direct com www Testo Direct com ...

Page 7: ...ent cycle in a measurement series can be found Manual saving of a single measurement by pressing the save button Automatic saving program has been set up Saving is activated by pressing the button Symbol for reading contents of memory on display Symbol for deleting memory contents If this symbol appears the printing function is activated The symbol flashes while data is being transmitted You can p...

Page 8: ...omatic saving depending on setting 5 Print readings 2 Measurement functions HOLD Freeze values MAX Display maximum values MIN Display minimum values MEAN Multi point mean calculation Update values Calculate mean Reset mean MEAN Timed mean calculation Start and end Calculate mean Reset mean E ES SC C L Lo oc ca at ti io on n Save manually Print readings 3 Location selection 1 Selection of 99 locati...

Page 9: ...t selection Lux fc 6 Instrument configuration Factory reset MAN Activate manual saving AUTO Program automatic saving Select no of meas cycles Set measurement rate OUT Read memory contents on display with printing option CLEAR Clear memory Note Everything in the memory is deleted Note Flashing display is activated and is confirmed by pressing 1 888 610 7664 info Testo Direct com www Testo Direct co...

Page 10: ...he following functions during mea suring at the touch of a button Save readings Manual or automatic saving is determined by the save setting Chapter 5 Print readings I 0 V BAT 7 5 1 Location Museum Location Line 1 Line 2 824 M 05 MAN N 0017 Lux Museum 1 888 610 7664 info Testo Direct com www Testo Direct com ...

Page 11: ...gs since the measurement began 824 HLD Museum M 05 MAN N 0017 Lux 824 Lux Museum M 05 MAN N 0017 x10 856 MAX Museum M 05 MAN N 0017 Lux 612 MIN Museum M 05 MAN N 0017 Lux Hold Max Min Mean Hold Max Min Mean Hold Max Min Mean Hold Max Min Mean Save reading Print reading Save maximum reading Print maximum reading Save reading Print reading ESC Location 1 888 610 7664 info Testo Direct com www Testo ...

Page 12: ...umber of values logged Mean Current measurement 823 Museum M 05 MAN N 0017 Lux Hold Max Min Mean Hold Max Min Mean Hold Max Min Mean New mean calculation possible ESC Location Note on saved or printed logs 1 The log of a multi point mean calculation contains single values max value min value and mean value 2 The log of a timed mean calculation contains max value min value and mean value Return to ...

Page 13: ...int mean New mean calculation is possible Hold Max Min Mean Current measurement 824 Museum M 05 MAN N 0017 Lux Hold Max Min Mean Hold Max Min Mean Hold Max Min Mean Hold Max Min Mean 824 00 00 S ST TA AR RT T M 05 MAN N 0017 Lux Note on saved or printed logs 1 The log of a multi point mean calculation contains single values max value min value and mean value 2 The log of a timed mean calculation c...

Page 14: ... load your own location names 8 characters onto your instrument using PC software Selection of location Confirm selection and return to measurement 14 Location selection 824 Museum M 05 MAN N 0017 Lux 1 Current measurement 0 01 1 Location Museum E ES SC C L Lo oc ca at ti io on n or From now on all measured data which is saved or printed is linked to the selected location or product names Return t...

Page 15: ...p line The function is activated with Press the button to get to the memory settings mode Confirm flashing save symbol via button 1 Current measurement MEMORY MAN AUTO O OU UT T CLEAR M MA AN N AUTO A AU UT TO O OUT C CL LE EA AR R MA MEMORY MEMORY MEMORY ESC Location ESC Location Parameter settings 824 Museum M 05 MAN N 0017 Lux 1 888 610 7664 info Testo Direct com www Testo Direct com ...

Page 16: ...et value by pressing 2 Cycle Select interval in which the measured values are to be saved The blinking position can be changed using Confirm set value by pressing Automatic saving is started by pressing The symbol flashes until the programmed measurement series is accepted A AU UT TO O OUT MEMORY MAN M MA AN N AUTO MORY 660 0 AUTO CYCLE N ESC Location 00 01 30 0 AUTO CYCLE h min sec M 06 N 0017 A ...

Page 17: ...ctivates a printout of the selected measurement log Press to display the values measured Cancel printout by pressing the button 4 Select the measurement cycle only possible if a measurement log consists of a measurement series Confirm the measurement log selected above by pressing The N 0x counter flashes in the display Select the measurement cycle by pressing activates a printout of the selected ...

Page 18: ...access the memory settings mode Confirm flashing save symbol via the button i Smith Ltd 31 08 1998 11 26 05 Museum M 01 31 08 1998 11 20 05 1 725 Lux Info Printout of manual saving Heading Date of printout Location Log Start of saving Measuring cycle Printout of automatic saving Smith Ltd 31 08 1998 11 27 32 Museum M 05 31 08 1998 11 26 59 cycle 00 00 02 1 Lux 01 653 02 652 03 650 04 690 05 705 06...

Page 19: ... of an activated function cycle 1 min the instrument switches itself on at the measurement time and switches off again This also occurs if the instrument is switched off via the button after the saving program has been activated BAT Battery operation with 9 V block battery Alkali manganese IEC 6LR61 AKKU Rechargeable battery operation with Testo rechargeable battery Part no 0515 0025 Type Ni MH IE...

Page 20: ...hes to the display of the cur rent measurement Unit Lux fc Current measurement 20 Instrument configuration Select unit Unit selection Factory reset Power supply Set date time 01 01 199 98 8 DATE 11 25 TIME Setting date The flashing position in the display can be set scroll next position The blinking position in the display can be set scroll next position Setting the time 1 888 610 7664 info Testo ...

Page 21: ... range has not The measured values are been reached outside the allowed measuring range Switch resolution Connection to the probe Please contact a Testo has been interrupted service point Error message Cause Remedy 1 888 610 7664 info Testo Direct com www Testo Direct com ...

Page 22: ... 000 Lux 10 Lux 0 to 3 200 fc 0 1 fc 0 to 10 000 fc 1 fc Display 2 line LCD and 2 matrix lines Battery 9 V IEC 6F22 lifetime 50 h Al Mn Battery Automatically check in 4 stages Operating temperature 0 to 50 C Storage temperature 20 to 70 C Dimensions 220 x 68 x 50 instrument Weight 500 g incl packaging Warranty Instrument 2 years Probes 12 months 1 888 610 7664 info Testo Direct com www Testo Direc...

Page 23: ...e analysis and graphics function data analysis trend curve 0554 0830 0554 0547 ComSoft 3 Professional with data management 0516 0445 0409 0178 Transport case Calibration point 1000 Lux 0520 0010 ISO calibration certificate Ordering d data f for testo 545 With bench stand and belt clip protects measuring instrument from impact dirt RS232 cable Connects measuring instrument PC for data transmission ...

Page 24: ...0973 5450 en 03 V01 70 1 888 610 7664 info Testo Direct com www Testo Direct com ...

Reviews: