7 Maintaining the product
33
7
Maintaining the product
Changing the battery/rechargeable battery
To avoid the loss of data, it is imperative that you switch
the instrument off when changing the
battery/rechargeable battery and replace the battery in
< 10 min.
1.
Open the battery compartment on the rear of the instrument.
2.
Remove the empty monobloc/rechargeable battery.
3.
Insert new empty monobloc/rechargeable battery.
> Observe the polarity!
4.
Close the battery compartment.
Charging rechargeable battery
Incorrect charging of batteries!
Danger of explosions.
Only start the charging process if a rechargeable
battery is in the instrument and the battery type is set to
RechBatt
.
1.
Check if a rechargeable battery is in the instrument.
2.
Check if the battery type is set to
RechBatt
.
3.
Connect the connector of the mains unit to the
12 V DC
socket
of the instrument.
4.
Connect the mains plug to the mains socket.
- A message box is displayed asking whether the rechargeable
battery should be charged.
5.
Select
Yes
with
and confirm with
.
The charging process begins automatically.
flashes during the
charging process and the current battery voltage is displayed.
Cleaning the instrument
>
If the housing of the instrument is dirty, clean it with a damp
cloth.
Do not use any aggressive cleaning agents or solvents! Weak
household cleaning agents and soap suds may be used.