6 Using the product
17
•
Fast
: 25 measurements are automatically recorded per
second by pressing
in the measurement view. A fast
measurement is only possible with pressure probes or an
internal pressure sensor. Only 1 channel can be evaluated
and the following sequence applies: external pressure
sensor before internal pressure sensor, channel 2 before
channel 1.
Configuration
Set
Auto
measurement program
(if the function
Manual
or
Automatic
was selected under
Man/Auto
):
1.
Set measuring rate using
or
. Hold the key down to go
forward/back quickly.
2.
Confirm the selection with
.
3.
Set number of measurements using
or
. Hold the key
down to go forward/back quickly.
- The duration of the measurement series is displayed in the
upper line for your information.
4.
Confirm the selection with
.
Set
Fast
measurement program (if the function
Fast
was selected
under
Man/Auto
):
1.
Select number of measurements using
or
. Hold the
key down to go forward/back quickly.
2.
Confirm the selection with
.
The protocols recorded for a location such as reading or other
available parameters (density, temperature, humidity, pressure,
section, offset factor, Pitot tube factor) can be printed.
- If no protocol is recorded,
Error
is shown in the display.
1.
Use
or
to select the protocol. Hold the key down to
go forward/back quickly.
2.
Confirm the selection with
.
- The data is sent to the printer via the infrared interface.
flashes during the data transfer.