Control and calibration options

The control and calibration set (Order no.: 0554.0660)

is used to control and calibrate humidity probes. The

set consists of two special containers. 

With the aid of different saline solutions, air mixtures

with defined relative humidities are produced (for

additional information see the Instruction Manual

“Control and calibration set for humidity sensors”).

As part of the accreditation as a calibration laboratory

for the parameters 

“Relative humidity”

and

“Dew point”

the values for the relative humidity over

the saturated lithium chloride (LiCl) and sodium

chloride (NaCl) solutions were newly specified at 

testo

.

These improved values also apply to all of the control

and calibration sets already delivered. Ignore the

values specified up to now.

9

LiCl 11.3%RH, NaCl 75.3%RH

at rated tempe25 °C

Temperature dependencies

Lithium chloride

Temperature

LiCl

Average

Tolerance 

10 °C

11.29 %

± 0.41 %

15 °C

11.30 %

± 0.35 %

20 °C

11.31 %

± 0.31 %

25 °C

11.30 %

± 0.27 %

30 °C

11.28 %

± 0.24 %

Sodium chloride

Temperature

NaCl

Average

Tolerance

10 °C

75.67 %

± 0.22 %

15 °C

75.61 %

± 0.18 %

20 °C

75.47 %

± 0.14 %

25 °C

75.29 %

± 0.12 %

30 °C

75.09 %

± 0.11 %

Summary of Contents for 0636.9835

Page 1: ...Pressure Dew Point Probe 0636 9835 0636 9836 Application Information en 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Page 2: ...ressure dew point probe 6 Measuring 7 Maintenance 8 Control and calibration options 9 Technical data 12 Warranty 12 Ordering data 13 Appendix A 14 Necessary pressure dew point values tpd 14 Appendix B...

Page 3: ...Probe description 3 PTFE sintered cap Humidity sensor Temperature sensor Standard plug in nipple Humidity temperature probe Internal thread Measuring chamber Protective sleeve valve...

Page 4: ...e 100 RH This value is an important criteria for the perfect running of the compressed air plant The probe is connected to the compressed air system via a standard plug in connector G1 4 internal thre...

Page 5: ...sure range Observe the measuring range of the sensor Overheating can destroy the probe Observe the permissible storage and transport temperature as well as the permissible operating temperature e g pr...

Page 6: ...air Connect the probe to the instrument In order to obtain a real measuring value let compressed air escape from the outlet for about 10 seconds Connect the humidity probe to the compressed air plant...

Page 7: ...rage of the humidity probe in areas with temperature and humidity values which differ considerably from those of the compressed air being measured and dirty sensors or sintered caps increase the respo...

Page 8: ...e sensor in the alcohol rinse in distilled water and then leave to dry in the air Put the PTFE sintered cap back on and screw tightly Faulty sensors can be replaced please contact our service departme...

Page 9: ...ers Relative humidity and Dew point the values for the relative humidity over the saturated lithium chloride LiCl and sodium chloride NaCl solutions were newly specified at testo These improved values...

Page 10: ...H 2 RH Screw out the probe from the LiCl test container Screw the probe into the NaCl 75 3 test container 15 minute adjustment time Control on the hand held instrument display Calibrate if necessary b...

Page 11: ...tion at least 15 minutes The probe 0636 9836 is already optimized in the factory for humidity measurement in compressed air systems How to carry out a fine calibration Screw off the protective sleeve...

Page 12: ...Resolution 0 1 RH 0 1 C Temperature range Probe 20 to 50 C Instrument 0 to 40 C Control point 11 3 to 75 3 RH Measuring air flow rate at 6 0 bar overpressure 1 0 l min factory set at delivery relatin...

Page 13: ...robe 0636 9836 Measuring chamber Instruction Manual Measuring chamber 0554 3303 Humidity sensor 0420 0023 Temperature sensor 0420 1242 Humidity control and calibration set 0554 0660 11 3 RH 75 3 RH In...

Page 14: ...alues tpd Factory air internal pipes 10 C to 10 C Paint spraying 10 C to 25 C Instrument air 10 C to 40 C Sand blasting machine 5 C to 0 C Pneumatic tools 5 C to 25 C Pneumatic conveyor belt 5 C to 60...

Page 15: ...6 5 6 0 Drucktaupunkt C td This graph gives information on the change in the pressure dew point at pressure loss A drop in working pressure from 8 bar to 6 bar is used as an example here In this case...

Reviews: