Warranty
Pressure dew point probe ................1 year
Measuring chamber .........................1 year
If the instrument is opened, improperly handled or if force is applied, the
warranty will be no longer valid.
12
Technical data
Sensor:
............................................
testo
®
ceramic sensor
Measuring range:
...........................0 to 100 %RH,
Pressure dew point t
pd
: .....................-60 to +50 °C t
pd
Pressure range:
..............................Overpressure to 15 bar
Accuracy, temperature:*
................±0.4 °C
Resolution:
.....................................0.1 %RH / 0.1 °C
Temperature range:
.......................Probe: -20 to +50 °C
Instrument: 0 to +40 °C
Control point:
.................................11.3 to 75.3%RH
Measuring air flow rate
at 6.0 bar overpressure: ................
1,0 l/min factory set at
........................................................delivery (relating to 0 bar,
........................................................+20°C)
Response time:
..............................1 to 5 min
typically 2 min
Storage/transport temperature:
....-20 to +70°C
Accuracy:
* Accuracy referred to a rated temperature of +25 °C.
The Technical data applies in connection with the
testo x35
measuring instrument.
5
0
-5
-10
-20
-30
-40
±0.9
±1
±2
±3
±4
-
-
±0.8
±0.8
±0.9
±1
±2
±3
±4
Accuracy
in °C t
pd
0636.9835
0636.9836
Pressure
dew point
in °C