![TESTO AG 400 Instruction Manual Download Page 27](http://html.mh-extra.com/html/testo-ag/400/400_instruction-manual_1092560027.webp)
53
52
Temperature
Pt100:
measuring range
-200 to +800 °C
Resolution
- 99.9 to 300 °C: 0.01 °C
Remaining range: 0.1 °C
Accuracy
±0.1 °C (-49.9 to +99.9 °C)
±0.4 °C(-99.9 to -50 and +100 to 199.9 °C)
±1.0 °C (-200 to -100 and +200 to 800 °C)
NiCr-Ni:
measuring range
-200 to +1370 °C
Resolution
0.1 °C
Accuracy
±0.4 °C (-100 to +200 °C)
±1 °C (rem. range)
PtRh-Pt:
measuring range
0 to +1760 °C
Resolution
1 °C
Accuracy
±1.0 °C
FeCu-Ni:
measuring range
-200 to +1000 °C
Resolution
0.1 °C
Accuracy
±0.4 °C (-150 to +150 °C)
±1 °C (rem. range)
Cu-CuNi:
measuring range
-40 to +350 °C
(Type T)
Resolution
0.1 °C
Accuracy
±0.4 °C (-40 to +200 °C)
±1 °C (rem. range)
NTC:
measuring range
-40 to 150 °C
Resolution
0.1 °C
Accuracy
±0.2 °C (-10 to 50 °C)
±0.4 °C (rem. range)
Humidity measurement
Measuring range
0 to 100 %RH
Resolution
0.1 %RH
Accuracy
See probe data
Pressure measurement
Measuring range
Resolution
Accuracy
±100 hPa
0.01 hPa
±0.1 hPa
±10 hPa
0.001 hPa
±0.01 hPa
2000 hPa
1 hPa
±2 hPa
10 bar
0.01 bar
±0.2 % of E type
30 bar
0.01 bar
±0.2 % of E type
rpm measurement
Measuring range:
20 to 20 000 rpm
Resolution:
1 rpm
Accuracy:
±1 digit
Velocity measurement
Vane probe
: measuring range
0 to 60 m/s
Resolution for ø 60/100 mm
0.01 m/s
Resolution for remaining probes
0.
m/s
Accuracy
See probe data
Therm. probes
.: measuring range
0 to 20 m/s
Resolution 0.01
m/s
Accuracy ±0.01 m/s (0 to 1.99 m/s)
±0.02 m/s (2 to 4.9 m/s)
±0.04 m/s (5 to 20 m/s)
Current/voltage measurement
(with current/voltage cable 0554.0007)
Measuring range:
0 to 20 mA
Resolution:
0.01 mA
Accuracy:
±0.04 mA
Measuring range:
0 to ±10 V
Resolution:
0.01 V
Accuracy:
±0.01 V
Measuring range:
0 to ±1 V
Resolution:
1 mV
Accuracy:
±1 mV
Memory space
Standard 45 000 measured values
Upgrade to 500 000 measured values
Power supply
•AlMn LR06 battery/
rechargeable battery (type: round cell)
• or via 8 V mains unit
Battery lifetime
Cont. operation w. 2 T/C probes: 18 h
Operating temp.
0 to +50 °C (short-term 0 to +60°C)
Storage/transport temperature
-25 to +60 °C
Connection
DIN 8 pin plug
Weight
500 g (incl. rechargeable battery)
Other features
• Automatic recognition of all
connected probes
Technical data
Measuring data
Update on disk
testo 650
and
testo 950
can be upgraded to have the full functions of
testo 400
.
We will
gradually introduce branch-specific and user-specific software updates for the instruments.
Please ask your Testo service department or Testo dealer.
A simple instrument update e. g. firmware V1.20 to V1.21 can be performed by you.
Remove the lithium battery, batteries/rechargeable batteries and mains unit from the measuring
instrument. Reconnect the mains unit. At the same time keep the
button on the instrument
pressed. The following message appears :
Connect your measuring instrument with an interface lead to your PC. Place the disk in disk
drive A:. Change to DOS input command (C:\). Type “A:\ <Enter> and then “Update” <Enter>.
All you have to do now is select the interface to which your Testo instrument is connected. A
progress bar indicates how far the update is...
Once the update has been installed successfully, switch off the instrument and replace the
batteries/rechargeable batteries.
Flash-Kernel
Active
No Connection
Notes:
The given times refer to batteries/rechargeables with the capacities mentioned
opposite. This standard data is supplied by the manufacturers. The values are
subject to fabrication spreads - the storage times and storage temperatures
also play a role.
Rechargeable batteries have to be charged and discharged regularly in order to
retain their rated capacity. After not using the almost empty batteries they will
only have a fraction of their capacity available..
If the system has not been used for a longer period of time (e.g. more than 4
weeks) remove all the batteries/rechargeable batteries from the instrument.
Rechargeable batteries should be stored in a charged condition separately.
Power supply