Store programming with the left function button.
The programmed memory function cannot be activated without “Save”.
Save:
activates the input programming, the left function button is automatically allocated with
“Start”, if necessary.
Erase:
programming is erased.
When storing programs are activated you will be informed via the display (in the top
left function field) about the progress in storing.
1. Program is active and is waiting for the starting criterium to be fulfilled:
Low limit:
Ð
High limit:
Ï
Manual:
Time start:
2. The program is running and is storing values.
appears along with a symbol from item 1.
When this symbol disappears, the program has finished and can be started again if
necessary.
PRINT - I
NSTRUMENT DATA
and
-
C
ONFIGURATIONS
produce a printout
which can be used for checking purposes.
Automatic storing
PROGRAMME
-
E
ND
- Memory
Full
stops the program if the complete memory is full
.
Wrap Around
Stores until storing has been interrupted by the the “Stop” function
button (appears automatically in the function bar once the
program has started). If the memory is full, the values stored at
the start are overwritten. Therefore you always have the last
possible values in the memory.
No. of Values
ends the program if the specified number of values in the memory
has been reached.
Date/Time
is only available with programmed starting criterium,
date/time.
The following sub-menus should be checked or set before storing:
PROGRAMME - S
TART
- Lower limit/Upper limit
A probe must be connected via which the program can be
started . Auto-OFF to “OFF”. Input the required limit values.
Note:
Do not switch off the instrument after programming
otherwise the starting criterium cannot be monitored.
- Manual
Start the program on location by pressing the function button.
The button is automatically allocated with “Start”.
- Date/Time
Does the program start at the specified time ?
(With measuring cycles greater than 2 min the instrument
switches off if Auto-Off was activated
→
power-save mode. It is
activated to start measuring).
PROGRAMME - C
YCLE TIME
:
determines the interval between two measurements. The time
unit can be selected via the function buttons: s, min or h.
Note:
The starting time must lie in the future.
Automatic storing
Note:
Some probes need a certain “Waking up” time. It is not possible to measure during this
time. A counter going backwards is blended in. This is taken into consideration when
automatic storage out of the sleep mode takes place (Auto OFF).
Probes which have to be zeroised after being switched on should be used for longterm
logging in continuous operation, mode Auto-OFF (OFF).