![TESTING 1.0330 Operating Manual Download Page 19](http://html1.mh-extra.com/html/testing/1-0330/1-0330_operating-manual_1092044019.webp)
Cabinet for Moist Storage
of Mortar Prisms in Triple Moulds
1.0330
19
TROUBLE
CAUSE
SOLUTION
The humidifier is producing too much
moisture.
The humidifying pipes are not in their
correct positions.
There is no air circulation. The internal
fans are defective.
The controller settings are not correct.
Get in touch with the manufacturer.
Get in touch with the manufacturer.
Replace the circulation fans.
The interior of the Cabi-
net is too dry.
The water supply is not functioning
properly.
The humidifier is not switched on, or it
switched itself off.
If using a water tank
: refill with dis-
tilled water.
Check to make sure the water supply
is properly functioning. Get in touch
with the manufacturer.
The interior of the Cabi-
net is too warm.
The cooling system is defective. The
controller or the sensors are defective.
The controller settings are not correct.
The fans are defective. There is no
proper air circulation.
Get in touch with the manufacturer.
The interior of the Cabi-
net is too cold.
The heating system is defective. The
controller or the sensors are defective.
The controller settings are not correct.
The fans are defective. There is no
proper air circulation.
Get in touch with the manufacturer.
The controllers do not
show a measured
value, although the
main switch is ON.
Excessive voltage. Short circuit.
The miniature circuit breakers have
been turned OFF.
Take off the front panel of the Cabinet
and turn ON the miniature circuit
breakers in the electrical box. Find out
why the circuit breakers tripped. Get
in touch with the manufacturer.
6.
Placing the system out of operation
If the Cabinet should be placed out of operation for a lengthy period of time, it must be
disconnected from the mains power supply. Then, during this time, conduct all mainte-
nance on the system that is required. Drain or remove the water from all storage tanks.
Cover the Cabinet to protect it from dust. Keep the door open in order to prevent unpleas-
ant smells.
If the humidifier is not used over longer periods of time, the water must be drained.
7.
Scrapping the system
If the Cabinet cannot be further used, we recommend scrapping it as follows:
Disconnect the equipment from the mains power
Drain all the water out of the system
Dispose of the refrigerant in accordance with currently valid laws.
Take the system apart and scrap it in accordance with currently valid laws.