Laboratory mortar mixer
1.0205
16
6. Preparations for commissioning
6.1 Installation location/ambient conditions
Caution
The operation of the mortar mixer is only permissible in
dry spaces!
The following limit values apply:
Ambient temperature
DIN EN 60204-1, 4.4.2
+5
C up to +40
C
Humidity
DIN EN 60204-1, 4.4.3
30% up to 95%
Note
Place the mortar mixer on a stable, level surface.
6.2 Electrical connection
A protective connection socket
110-240 volts 1Ph+1N+1PE, 50-60 Hz
must be used for
the mains connection of the mortar mixer.
The fuse should be fused with a maximum of 16 A. A residual current circuit breaker is
recommended.
Insert the Schuko plug, the mains cable of the mortar mixer, into the mains connection
socket.
Caution
Safety advice!
Work on electrical equipment may only be carried out by qualified
personnel!