Chapter 8 – Maintenance
TestEquity 123H Temperature/Humidity Chamber
Page 8-19
Series F4 Setup Parameters
System
Main Page\Go to Setup\System
GSB 1 Source
Input 1
Guar. Soak Band 1
3.0
°
C (5.4 if °F)
GSB 2 Source
Input 2
Guar. Soak Band 2
5.0 %
Current Time
(local time)
Current Date
(local date)
PID Units
US
°
F or
°
C
°
C
Show
°
F or
°
C
Yes, Upper Display
Ch1 Autotune SP
90 %
Ch2 Autotune SP
90 %
Input 1 Fail
0 %
Input 2 Fail
0 %
Open Loop Ch1
Off
Open Loop Ch2
Off
Power Out Time
10 sec
Power-Out Action
Continue
Analog Input
Main Page\Go to Setup\Analog Input 1
Sensor
Thermocouple
Type
T
Choose Decimal
0.0
SP Low Limit
-75.0
°
C (-103.0 if °F)
SP High Limit
175.0
°
C (347.0 if °F)
Calibration Offset
0.0
°
C
Filter Time
1.0 sec
Error Latch
Self-Clear
Main Page\Go to Setup\Analog Input 2
Sensor
Process
Type
0 to 20mA
Units
% rh
Decimal
0.0
Scale Low
0.0 %
Scale High
100.0 %
Scaling
Normal Scaling
SP Low Limit
0.0 %
SP High Limit
98.0 %
Calibration Offset
0.0
Filter Time
1.0 sec
Error Latch
Self-Clear
Main Page\Go to Setup\Analog Input 3
Sensor
Thermocouple
Type
J
Decimal
0
SP Low Limit
0
°
C
(
32 if °F)
SP High Limit
815
°
C (1500 if °F)
Calibration Offset
0
°
C
Filter Time
1.0 sec
Error Latch
Latch
Cascade
No Cascade
Digital Input Menu
Main Page\Go to Setup\Digital Input (1-4)
Name
No
Function
Off
Control Output
Main Page\Go to Setup\
Control Output 1A
Function
Heat
Choose Cycle Time Fixed Time
Enter Cycle Time
3.0 sec
Hi Power Limit
100 %
Low Power Limit
0 %
Main Page\Go to Setup\
Control Output 1B
Function
Cool
Choose Cycle Time Fixed Time
Cycle Time
6.0 sec
Hi Power Limit
100 %
Low Power Limit
0 %
Main Page\Go to Setup\
Control Output 2A
Function
Heat
Choose Cycle Time Fixed Time
Cycle Time
2.0 sec
Hi Power Limit
100 %
Low Power Limit
0 %
Main Page\Go to Setup\
Control Output 2B
Function
Cool
Choose Cycle Time Fixed Time
Cycle Time
10.0 sec
Hi Power Limit
100 %
Low Power Limit
0 %
Alarm Output
Main Page\Go to Setup\
Alarm Output 1
Name
HUM OVRTMP
Alarm Type
Process
Alarm Source
Input 3
Latches
Self-Clears
Silencing
No
Alarm Hysteresis
25
°
C (45 if °F)
Alarm Sides
High
Alarm Logic
Open on Alarm
Show Message
Yes on Main Page
Main Page\Go to Setup\
Alarm Output 2
Name
HUM RANGE
Alarm Type
Process
Alarm Source
Input 1
Latches
Self-Clears
Silencing
No
Alarm Hysteresis
2.0
°
C (3.6 if °F)
Alarm Sides
Both
Alarm Logic
Open on Alarm
Show Message
No
Retransmit Output
(if retransmit option is ordered)
Main Page\Go to Setup\
Retransmit Output 1
Retransmit Source Input 1
Analog Range
0-5V
Low Scale
-75.0°C (-103.0 if °F)
High Scale
175.0°C (347.0 if °F)
Scale Offset
0.0
Main Page\Go to Setup\
Retransmit Output 2
Retransmit Source Input 2
Analog Range
0-5V
Low Scale
0.0
High Scale
100.0
Scale Offset
0
Digital Output
Main Page\Go to Setup\
Digital Output 1
Name
Yes
Name Digital Out. CONDITION
Function
Event Output
Main Page\Go to Setup\
Digital Output 2
Name
Yes
Name Digital Out. HUMIDITY
Function
Event Output
Main Page\Go to Setup\
Digital Output 3
Name
Yes
Name Digital Out. PURGE
Function
Event Output
Main Page\Go to Setup\
Digital Output (4-7)
Name
No
Function
Event Output
Main Page\Go to Setup\
Digital Output 8
Name
No
Function
Compressor
Comp. On % Pwr. -2%
Comp. Off % Pwr. 2%
Comp. Off Delay
60 sec
Comp. On Delay
10 sec
Communications Menu
Main Page\Go to Setup\
Communications
Baud Rate
9600
Address
1
Custom Main Page
Main Page\Go to Setup\
Custom Main Page
P1
Input 2
P2
Current File
P3
Current Step
P4
Set Point 1
P5
Set Point 2
P6
Step Type
P7
Target SP1
P8
Target SP2
P9
Wait For Status
P10
Jump Count
P11
Time Remaining
P12
Digital Outs
P13
% Power 1
P14
% Power 2
P15
Time
P16
Date
Process Display Menu
Main Page\Go to Setup\
Process Display
Choose
Input 1 only
Static Message Menu
Main Page\Go to Setup\
Static Message
Message 1 TESTEQUITY 123H
Summary of Contents for 123H
Page 2: ......
Page 6: ......
Page 10: ......
Page 20: ......
Page 44: ......
Page 48: ......
Page 72: ......
Page 74: ......
Page 75: ...Chapter 10 Drawings TestEquity 123H Temperature Humidity Chamber Page 10 1 Chapter 10 Drawings...
Page 76: ......
Page 77: ...Electrical Subpanel Component Location Purge Subpanel Main Subpanel...