Chapter 7 – Specifications
TestEquity 123H Temperature/Humidity Chamber
Page 7-1
Chapter 7 – Specifications
Model 123H Chamber Specifications
Temperature Range
–68°C to +175°C
Control Tolerance
±0.5°C (Measured at the control sensor after stabilization)
Uniformity
±0.5°C (Variations throughout the chamber after stabilization)
Standard Humidity Range
10% to 95%, in the temperature range of 10°C to 65°C limited by a 6°C dewpoint;
90% over 65°C up to 85°C. See chart on page 4-6.
Control Tolerance
±3%RH (Measured at the control sensor after stabilization)
Live Load Capacity @
+23°C
0°C
–40°C
–55°C
–65°C
500 W
400 W
225 W
175 W
100 W
(Live load capacity is for non-humidity mode)
Cool Down Transition Time*
End Temp
Start Temp to
+23°C
0°C
–40°C
–55°C
–65°C
–68°C
+23°C
----
6 min
31 min
51 min
68 min
ultimate
+50°C
7 min
16 min
44 min
62 min
82 min
ultimate
+85°C
18 min
28 min
54 min
82 min
101 min
ultimate
+150°C
43 min
54 min
87 min
108 min
129 min
ultimate
Heat Up Transition Time*
End Temp
Start Temp to
+23°C
+50°C
+85°C
+125°C
+150°C
+175°C
+23°C
-----
2 min
12 min
24 min
33 min
ultimate
0°C
3 min
9 min
18 min
32 min
41 min
ultimate
–40°C
9 min
16 min
25 min
39 min
48 min
ultimate
–55°C
14 min
18 min
31 min
44 min
53 min
ultimate
–65°C
15 min
21 min
32 min
45 min
54 min
ultimate
*Note:
Transition times are measured after a 2-hour soak at the start temperature with an empty chamber.
Input Power
120 V nominal (110 to 126 VAC), 60 Hz, 1 PH
Max Current Draw 18 A, Recommended Minimum Service 20 A (dedicated line)
Workspace Dimensions
18" W x 16.5" H x 13.5" D (2.3 cubic feet)
Outside Dimensions
26" W x 63" H x 36.5" D (nominal). Vent tube adds 3" to height.
Min. Installed Clearance
6" from the left and right side, 12" from the rear
Access Ports
4" Port on left and right side (two total), Supplied with foam plugs
Sound Level
62 dBA (A-weighted, measured 36" from the front)
Weight
750 pounds
NOTE:
Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed
for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased
cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or
below 16°C (60°F) ambient is not recommended.
Summary of Contents for 123H
Page 2: ......
Page 6: ......
Page 10: ......
Page 20: ......
Page 44: ......
Page 48: ......
Page 72: ......
Page 74: ......
Page 75: ...Chapter 10 Drawings TestEquity 123H Temperature Humidity Chamber Page 10 1 Chapter 10 Drawings...
Page 76: ......
Page 77: ...Electrical Subpanel Component Location Purge Subpanel Main Subpanel...